X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Technical Note: Characterization of x-ray beam profiles for a fluoroscopic system incorporating copper filtration
Kevin A Wunderle1, Andrew R Godley2, Zhilei Liu Shen3
1Department of Radiology, Cleveland Clinic, 9500 Euclid Ave, Cleveland, OH, 44195, USA.
Purpose:
The goal of this study was to investigate x-ray beam profiles at various water depths to characterize the two-dimensional x-ray dose distribution, allowing for off-axis and out-of-field radiation dose estimation for a wide range of x-ray beam spectra commonly encountered in fluoroscopically guided interventional procedures.
Methods:
A Siemens Artis interventional fluoroscope was operated in a service mode to generate a continuous x-ray beam at fixed x-ray beam spectra, defined by their kVp and the thickness of additional copper filtration. A PTW scanning water tank with a diode detector was used to measure the x-ray beam profiles at several depths in water at various fields of view and x-ray beam spectra, both parallel and perpendicular to the anode-cathode axis direction.
Results:
X-ray beam profiles, including out-of-field tails, were characterized for a wide range of beam qualities. The anode heel effect was pronounced even at depth, resulting in large dose variations across the x-ray field; this effect was even more definite at large fields of view, at higher kVps, and in the absence of additional copper filtration.
Conclusions:
This study investigated and characterized 2D radiation dose deposition in water from x-ray beam spectra commonly used by modern fluoroscopes in interventional procedures. This knowledge can be applied to manual dosimetry calculations or can be used to refine the accuracy of automated dose mapping tools or Monte Carlo simulations of the radiation dose to soft tissue within the x-ray field and to tissue adjacent to the primary beam. Additionally, this study illustrates a substantial reduction of the anode heel effect by using moderate amounts of additional copper filtration to harden the x-ray beam spectrum.
Related Concept Videos
Filtration and Urine Formation
Glomerular Filtration: Net Filtration Pressure
GBHP, with an average value of 55 mmHg, promotes filtration by pushing water and solutes through the filtration membrane. This is balanced by two opposing forces: CHP, a "back pressure" exerted against the filtration membrane by fluid already in the capsular space and renal...
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Filtration
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Glomerular Filtration
Components of the Filtration Membrane
The filtration process involves three key layers: the glomerular endothelial cells, the basement membrane, and the podocyte-formed filtration slits.