Value of spectral detector computed tomography for assessment of pancreatic lesions
Nada El Kayal1, Simon Lennartz2, Sandra Ekdawi3
1University of Cologne, Faculty of Medicine and University Hospital Cologne, Department of Diagnostic and Interventional Radiology, Cologne, Germany; Department of Diagnostic and Interventional Radiology, Alexandria University, Faculty of Medicine, Alexandria, Egypt.
Purpose:
Dual energy CT (DECT) can contribute to the diagnosis of benign and malignant pancreatic lesions. This study examined whether a novel, detector-based spectral CT scanner (SDCT) may improve subjective assessment of different types of pancreatic lesions and if various quantitative maps may improve lesion contrast and differentiation.
Materials And Methods:
61 consecutive patients who underwent clinical, contrast-agent enhanced, abdominal SDCT scans and showed pancreatic lesions of different origins were included. Subjective image analysis was performed by two readers who assessed image quality, lesion conspicuity and diagnostic confidence on 5-point Likert scales for conventional polyenergetic reconstructions (polyE), virtual monoenergetic images (monoE), virtual non-contrast images, iodine density, iodine overlay, and Z effective (Zeff) maps. Two readers acquired quantitative values from these maps ROI-based from which contrast-to-noise and lesion-to-parenchyma ratios were calculated.
Results:
MonoE images at low keV levels yielded highest Likert scores regarding lesion conspicuity and reader confidence; iodine overlays facilitated lesion delineation. Inter-observer agreement ranged between substantial and excellent (kappa values 0.73-0.81). Contrast-to-noise-ratios for low keV monoE images were significantly higher, compared to polyE images (e.g. monoE 40 keV p < 0.0001). Marked overlap between PDAC and miscellaneous non-PDAC lesions was present in various spectral reconstructions.
Conclusions:
In line with previous studies, monoE images at low keV levels and iodine overlay maps facilitated subjective lesion delineation which was substantiated by the quantitative analysis. Hence, spectral detector CT improves pancreatic lesion conspicuity, while its value for lesion differentiation needs to be further evaluated in larger study cohorts.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors