Current Generation Balloon-Expandable Transcatheter Valve Positioning Strategies During Aortic Valve-in-Valve
Matheus Simonato1, John Webb2, Sabine Bleiziffer3
1Division of Cardiac Surgery, Escola Paulista de Medicina - Universidade Federal de São Paulo, São Paulo, Brazil; Division of Cardiology, St. Paul's Hospital, Vancouver, British Columbia, Canada.
Objectives:
This study sought to evaluate SAPIEN 3 (S3) (Edwards Lifesciences, Irvine, California) positioning using different strategies.
Background:
Aortic valve-in-valve (ViV) is associated with high risk of elevated gradients.
Methods:
S3 aortic ViV procedures in stented bioprostheses were studied. Transcatheter heart valve (THV) positioning was analyzed in a centralized core lab blinded to clinical outcomes. A combined endpoint of severely elevated mean gradient (≥30 mm Hg) or pacemaker need was established. Two positioning strategies were compared: central marker method and top of S3 method. Optimal final depth was defined as S3 depth ≤20%.
Results:
A total of 113 patients met inclusion criteria and were analyzed (76.5 ± 9.7 years of age, 65.8% male, STS score 8 ± 7.6%). THVs had incomplete shortening in comparison to fully expanded valves (92 ± 3.4%), and expansion was more complete in optimal positioning cases compared with others (93.2 ± 2.7% vs. 91.5 ± 3.5%; p = 0.027). The central marker method demonstrated greater correlation with final implantation depth than the top of S3 method (R2 of 0.48 and 0.14; p < 0.001 and p = 0.001, respectively). The combined endpoint rate was 4.3% in the optimal (higher than 3 mm) implantation group, 12% in the intermediate group, and 50% in the low group (p < 0.001). There were no cases of THV embolization. In cases with central marker higher than 3 mm, 72.4% had optimal final depth. In those with central marker higher than 6 mm, 90% had optimal final depth.
Conclusions:
Optimal S3 positioning in aortic ViV is associated with better outcomes. Central marker positioning is more reliable than top of S3 positioning. Central marker bottom position should be 3 mm to 6 mm above the ring.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Predicting Reaction Outcomes