Related Experiment Video
Updated: Jan 20, 2026

Immunoglobulin G N-Glycan Analysis by Ultra-Performance Liquid Chromatography
Published on: January 18, 2020
Determination of Seven Manufacturing Impurities in FD&C Red No. 40 by Ultra-High-Performance Liquid Chromatography
Bhakti Petigara Harp1, Marianita Perez-Gonzalez1, Hector M Cordero-Nieves1
1U.S. Food and Drug Administration, Center for Food Safety and Applied Nutrition, Office of Cosmetics and Colors, College Park, MD 20740.
Background:
The U.S. Food and Drug Administration batch-certifies color additives to ensure that each lot meets published specifications for coloring food, drugs, and cosmetics.
Objective:
An ultra-high-performance LC (UHPLC) method was developed to determine seven manufacturing impurities in the monoazo color additive FD&C Red No. 40 (R40). The analytes consist of two intermediates, an impurity originating from one intermediate, a reaction by-product, and three subsidiary colors. The intermediates are 4-amino-5-methoxy-2-methylbenzenesulfonic acid [cresidine-p-sulfonic acid (CSA)] and 6-hydroxy-2-naphthalene sulfonic acid sodium salt (SS). The impurity originating from the intermediate SS is 6,6'-oxybis[2-naphthalenesulfonic acid] disodium salt. The reaction by-product is 4,4'-(diazoamino)bis[5-methoxy-2-methylbenzenesulfonic acid disodium salt. The subsidiary colors are sodium salts of CSA coupled with 2-naphthol-3,6-disulfonic acid, 2-naphthol-6,8-disulfonic acid, or 2-naphthol.
Methods:
Samples of R40 were dissolved in an ammonium acetate buffer modified to pH 9.2, filtered, and analyzed by UHPLC. Quantitation of the analytes was performed by calibration in the presence of the color additive matrix.
Results:
UHPLC validation studies showed linear calibration curves (R2 = 0.9999), good recovery (95-121%) and precision (RSDs = 1.0-6.3%), and LOQs ranging from 0.002 to 0.030%. Survey analyses of 31 samples from 11 manufacturers yielded results by the new UHPLC method and a previously used HPLC method that are consistent within experimental error.
Conclusions:
The new UHPLC method provides faster analysis time, improved separation, and similar sensitivity compared to the HPLC method.
Highlights:
An UHPLC method was developed and validated to determine seven manufacturing impurities in R40 submitted to the FDA for batch certification.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors
Steel Manufacturing
During this smelting process, limestone plays a crucial role by forming slag. Slag captures impurities within the molten iron, such...
Red Algae