High correlation between radiation dose estimates for 256-slice CT obtained by highly parallelized hybrid Monte Carlo
Peter Prinsen1, Sigal Trattner2, Jens Wiegert1
1Philips Research, Eindhoven, 5656AE, The Netherlands.
Purpose:
Accurate, patient-specific radiation dosimetry for CT scanning is critical to optimize radiation doses and balance dose against image quality. While Monte Carlo (MC) simulation is often used to estimate doses from CT, comparison of estimates to experimentally measured values is lacking for advanced CT scanners incorporating novel design features. We aimed to compare radiation dose estimates from MC simulation to doses measured in physical anthropomorphic phantoms using metal-oxide semiconductor field-effect transistors (MOSFETs) in a 256-slice CT scanner.
Methods:
Fifty MOSFETs were placed in organs within tissue-equivalent anthropomorphic adult and pediatric radiographic phantoms, which were scanned using a variety of chest, cardiac, abdomen, brain, and whole-body protocols on a 256-slice system. MC computations were performed on voxelized CT reconstructions of the phantoms using a highly parallel MC tool developed specifically for diagnostic X-ray energies and rapid computation. Doses were compared between MC estimates and physical measurements.
Results:
The average ratio of MOSFET to MC dose in the in-field region was close to 1 (range, 0.96-1.12; mean ± SD, 1.01 ± 0.04), indicating outstanding agreement between measured and simulated doses. The difference between measured and simulated doses tended to increase with distance from the in-field region. The error in the MC simulations due to the limited number of simulated photons was less than 1%. The errors in the MOSFET dose determinations in the in-field region for a single scan were mainly due to the calibration method and were typically about 6% (8% if the error in the reading of the ionization chamber that was used for the MOSFET calibration was included).
Conclusions:
Radiation dose estimation using a highly parallelized MC method is strongly correlated with experimental measurements in physical adult and infant anthropomorphic phantoms for a wide range of scans performed on a 256-slice CT scanner. Incorporation into CT scanners of radiation-dose distribution estimation, employing the scanner's reconstructed images of the patient, may offer the potential for accurate patient-specific CT dosimetry.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Field Effect Transistor
Metal-Semiconductor Junctions
Schottky Barriers
Schottky barriers arise when a metal with a work function (Φm) contacts a semiconductor with a different work function (Φs). Initially, electrons transfer until the Fermi levels of the metal and semiconductor align at equilibrium. For instance, if Φm > Φs, the semiconductor Fermi level is higher than the metal's before contact. The...
Biasing of Metal-Semiconductor Junctions
In Schottky junctions, where the semiconductor is n-type, applying a positive voltage to the metal relative to the semiconductor reduces its Fermi...
Estimation of the Physical Quantities
Biological Effects of Radiation