Impact of Spacing and Orientation on the Scar Threshold With a High-Density Grid Catheter
Masateru Takigawa1,2, Jatin Relan1,3, Takeshi Kitamura1
1IHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, Bordeaux University Hospital (CHU), Electrophysiology and Ablation Unit, University of Bordeaux, Centre de recherche Cardio-Thoracique de Bordeaux, France/Pessac-Bordeaux, France (M.T., J.R., T.K., C.A.M., S.K., R.M., G.C., K.V., G.M., A.F., N.T., M.W., F.B., A.L., J.D., T.P., A.D., N.D., X.P., J.M., J.N., M.M., F.C., B.Q., H.C., M. Hocini, M. Haïssaguerre, F.S., P.J.).
Background:
Multipolar catheters are increasingly used for high-density mapping. However, the threshold to define scar areas has not been well described for each configuration. We sought to elucidate the impact of bipolar spacing and orientation on the optimal threshold to match magnetic resonance imaging-defined scar.
Method:
The HD-Grid catheter uniquely allows for different spatially stable bipolar configurations to be tested. We analyzed the electrograms with settings of HD-16 (3 mm spacing in both along and across bipoles) and HD-32 (1 mm spacing in along bipoles and 3 mm spacing in across bipoles) and determined the optimal cutoff for scar detection in 6 infarcted sheep.
Results:
From 456 total acquisition sites (mean 76±12 per case), 14 750 points with the HD-16 and 32286 points with the HD-32 configuration for bipolar electrograms were analyzed. For bipolar voltages, the optimal cutoff value to detect the magnetic resonance imaging-defined scar based on the Youden's Index, and the area under the receiver operating characteristic curve (AUROC) differed depending on the spacing and orientation of bipoles; across 0.84 mV (AUROC, 0.920; 95% CI, 0.911-0.928), along 0.76 mV (AUROC, 0.903; 95% CI, 0.893-0.912), north-east direction 0.95 mV (AUROC, 0.923; 95% CI, 0.913-0.932), and south-east direction, 0.87 mV (AUROC, 0.906; 95% CI, 0.895-0.917) in HD-16; and across 0.83 mV (AUROC, 0.917; 95% CI, 0.911-0.924), along 0.46 mV (AUROC, 0.890; 95% CI, 0.883-0.897), north-east direction 0.89 mV (AUROC, 0.923; 95% CI, 0.917-0.929), and south-east direction 0.83 mV (AUROC, 0.913; 95% CI, 0.906-0.920) in HD-32. Significant differences in AUROC were seen between HD-16 along versus across (P=0.002), HD-16 north-east direction versus south-east direction (P=0.01), HD-32 north-east direction versus south-east direction (P<0.0001), and HD-16 along versus HD-32 along (P=0.006). The AUROC was significantly larger (P<0.01) when only the best points on each given site were selected for analysis, compared with when all points were used.
Conclusions:
Spacing and orientation of bipoles impacts the accuracy of scar detection. Optimal threshold specific to each bipolar configuration should be determined. Selecting one best voltage point among multiple points projected on the same surface is also critical on the Ensite-system to increase the accuracy of scar-mapping.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
08:20In Vitro Model of Human Cutaneous Hypertrophic Scarring using Macromolecular Crowding
Published on: May 1, 2020
Related Concept Videos
State Space Representation
Consider an RLC circuit, a...
Space Trusses
At the core of a space truss lies the fundamental unit known as the tetrahedron. This structure is composed of six members that form a three-dimensional shape...
Location and Orientation of the Heart
Impact of Groups on Groups
Transfer Function to State Space
In an RLC...
State Space to Transfer Function
The transformation process begins with the state-space representation, characterized by the state equation and the output equation. These equations are typically represented as: