Related Experiment Video
Updated: Jan 20, 2026

Brain Morphology of Cannabis Users With or Without Psychosis: A Pilot MRI Study
Published on: August 18, 2020
A pilot study of highly accelerated 3D MRI in the head and neck position verification for MR-guided radiotherapy
Yihang Zhou1, Oi Lei Wong1, Kin Yin Cheung1
1Medical Physics and Research Department, Hong Kong Sanatorium & Hospital, Hong Kong, China.
Background:
To evaluate the performance of a highly accelerated 3D MRI on inter-fractional positional measurement for MR-guided radiotherapy (MRgRT) in the head and neck (HN).
Methods:
Fourteen healthy volunteers received 159 scans on a 1.5 T MR-sim to simulate MRgRT fractions. MRI acquisition included a high-resolution (HQI-MRI, voxel-size =1.05×1.05×1.05 mm3, duration =5 min) and a highly-accelerated low-resolution (true-LQI-MRI, acceleration-factor =9, voxel-size =1.4×1.4×1.4 mm3, duration =86 s) T1w spin-echo sequence (TR/TE =420/7.2 ms). The first session HQI-MRI was used as the reference to mimic planning MRI. Other HQI-MRI was also retrospectively down-sampled in K-space and GRAPPA reconstructed to generate pseudo-LQI-MRI. Inter-sessional positional shift calculated from HQI-MRI, true-LQI-MRI and pseudo-LQI-MRI rigidly registering to the reference were analyzed and compared in the overall HN and the sub-regions of brain, nasopharynx, oropharynx and hypopharynx.
Results:
The calculated SD of systematic errors (Σ) from HQI-MRI/pseudo-LQI-MRI/true-LQI-MRI images for overall HN were 1.11/1.14/1.08, 0.28/0.26/0.29, 0.43/0.44/0.60, and 0.77/0.79/0.74 mm for translation in LR, AP, SI and 3D, respectively; The corresponding RMS of random errors (σ) were 0.97/0.98/0.96, 0.28/0.27/0.26, 0.77/0.77/0.72, and 0.85/0.87/0.85 mm. For all sub-regions, brain showed the smallest Σ and σ in 3D. Other sub-regions showed direction-dependent error patterns, but the positioning results were consistent, independent of the datasets used for registration.
Conclusions:
A highly-accelerated 3D-MRI could be used for MR-guided HN radiotherapy without compromising position verification accuracy.
Related Concept Videos
Arteries of the Head and Neck
The internal carotid arteries supply blood to the anterior portion of the cerebrum. They enter the...
Veins of Head and Neck
On the other hand, the vertebral veins, unlike their arterial counterparts, are not primarily responsible for brain drainage. Instead, they drain the cervical vertebrae, spinal cord, and some small...
Self-Evaluation: Self-Enhancement and Self-Verification
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Strategies of Self-Presentation II: Self-Verification
Pilot and Numeric Relaying