Characterization of scintillating fibers for use as positron detector in positron emission tomography
Vincent Turgeon1, Gustavo Kertzscher2, Liam Carroll3
1Medical Physics Unit, McGill University, Montreal, Quebec, Canada; Jewish General Hospital, Montreal, Quebec, Canada.
Purpose:
Manual and automatic blood sampling at different time intervals is considered the gold standard to determine the arterial input function (AIF) in dynamic positron emission tomography (PET). However, blood sampling is characterized by poor time resolution and is an invasive procedure. The aim of this study was to characterize the scintillating fibers used to develop a non-invasive positron detector.
Methods:
The detector consists of a scintillating fiber coupled at each end to transmission fiber-optic cables that are connected to photo multiplier tubes in a dual readout setup. The detector is designed to be wrapped around the wrist of the patient undergoing dynamic PET. The attenuation length and bending losses were measured with excitation from gamma radiation (137Cs) and ultraviolet (UV) light. The response to positron-emitting radio-tracers was evaluated with 18F and 11C.
Results:
The attenuation length for a 3.0 m and 1.5 m long scintillating fiber both coincides with the attenuation length given by the manufacturer when excited with the 137Cs source, but not with the UV source due to the differences in scintillation mechanisms. The bending losses are smaller than the measurement uncertainty for the 137Cs source irradiation, and increase when the bending radius decrease for the UV source irradiation. The signal-to-noise ratio for 18F and 11C solutions are 1.98 and 22.54 respectively. The measured decay constant of 11C agrees with its characteristic value.
Conclusion:
The performed measurements in the dual readout configuration suggest that scintillating fibers may be suitable to determine the AIF non-invasively.
More Related Videos
10:20Automation of a Positron-emission Tomography PET Radiotracer Synthesis Protocol for Clinical Production
Published on: October 26, 2018
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Emission Spectra
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...