In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Improved implant alignment accuracy with an accelerometer-based portable navigation system in medial unicompartmental
Yoshihito Suda1, Koji Takayama2, Kazunari Ishida1
1Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, 7-5-1 Kusunoki-cho, Chuo-ku, Kobe, 650-0017, Japan.
Purpose:
This study aimed to examine the accuracy of tibial implant alignment using an accelerometer-based portable navigation system in unicompartmental knee arthroplasty (UKA).
Methods:
This retrospective matched case-control study reviewed 51 UKAs performed using an accelerometer-based portable navigation system, matched with 51 UKAs performed using conventional extramedullary rods. Coronal alignment and posterior slope of the tibial implant were measured on postoperative radiographs, and differences from preoperative planning were examined. Outliers and accuracy of tibial implant alignment were compared between the portable navigation and conventional groups using Fisher's exact test and Mann-Whitney U test, respectively.
Results:
In the portable navigation group, 100% of the implants were aligned within 3.0° of both target coronal and sagittal implant alignment. In the conventional group, 76.5% and 88.2% of the implants were within 3.0° of both target coronal and sagittal implant alignment. Statistical analysis revealed that outliers of coronal and sagittal alignment were significantly less in the portable navigation group than in the conventional group (P < 0.05). In addition, the absolute value difference between postoperative measurement and preoperative planning of both coronal and sagittal alignment was significantly smaller in the portable navigation group than in the conventional group (P < 0.05).
Conclusion:
The portable navigation system improved the accuracy of tibial implant alignment in UKA. We found that 100% of the implants were aligned within 3.0° of both target coronal and sagittal implant alignment. The portable navigation system decreased the outliers of tibial coronal and sagittal alignment.
Level Of Evidence:
Retrospective case-control study, Level III.
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...