Independent assessment of source transit time for the BEBIG SagiNova® cobalt-60 high dose rate brachytherapy
Abolfazl Kanani1,2, Sareh Karbasi3, Mohammad Amin Mosleh-Shirazi4,5
1Ionizing and Non-Ionizing Radiation Protection Research Center, Paramedical School, Shiraz University of Medical Sciences, Shiraz, Iran.
Abstract:
Independent verification of transit time and the methodology employed in commercial high dose rate (HDR) afterloaders to compensate its effect is an important part of their commissioning and quality assurance. This study aimed to independently evaluate the Co-60 source transit time of the new BEBIG SagiNova® HDR afterloader unit by employing a dosimetric approach using a well-type ionization chamber. The source was placed at three dwell positions (DPs) to mimic a variety of clinical situations with different distances from the afterloader unit. The distances of the DPs to the afterloader were 129.37 cm, 124.50 cm and 118.57 cm. Plans were generated using the SagiPlan® treatment planning system to produce 3, 5, 10, 15, 20, 30, 40, 60 and 120 s dwell times (DTs). The residual transit times (following any possible system compensation) were assessed using the ESTRO-recommended approach of obtaining transit time compensation factors and another strategy established for teletherapy sources. The mean residual transit time depended on the distance between the afterloader and the DP, ranging from 0.43 to 1.10 s. The transit dose contribution was case-specific, ranging from 0.4% for a 60 s DT at the nearest DP to the afterloader up to 15.6% for a 3 s DT at the furthest DP from the unit. The results show that currently SagiNova® afterloader does not apply transit time compensation and suggest a 0.2-0.5 s compensation for each arrival and departure DP from/to the afterloader, depending on position in an 11 cm active length.
Related Concept Videos
Independent and Dependent Sources
Independent voltage or current sources supply a fixed amount of voltage or current, respectively, which is unaffected by other elements within the circuit. These are represented using specific symbols. Independent voltage sources are symbolized with polarities (+ and -), indicating the direction of the...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
The Integrated Rate Law: The Dependence of Concentration on Time
Cooperative Allosteric Transitions
Phase Transitions
Noncompartmental Analysis: Mean Transit, Absorption and Dissolution Time
One of the key parameters is the mean transit time (MTT), which refers to the total duration required for drug molecules to transit through the body. MTT is determined by calculating the ratio of the area under the moment curve to the area...