Rapid measurement and correction of spatiotemporal B0 field changes using FID navigators and a multi-channel
Tess E Wallace1,2, Onur Afacan1,2, Tobias Kober3,4,5
1Computational Radiology Laboratory, Department of Radiology, Boston Children's Hospital, Boston, Massachussetts.
Purpose:
To measure spatiotemporal B0 field changes in real time using FID navigators (FIDnavs) and to demonstrate the efficacy of retrospectively correcting high-resolution -weighted images using a novel FIDnav framework.
Methods:
A forward model of the complex FIDnav signals was generated by simulating the effect of changes in the underlying B0 inhomogeneity coefficients, with spatial encoding provided by a multi-channel reference image. Experiments were performed at 3T to assess the accuracy of B0 field estimates from FIDnavs acquired from a 64-channel head coil under different shim settings and in 5 volunteers performing deep-breathing and nose-touching tasks designed to modulate the B0 field. Second-order, in-plane spherical harmonic (SH) inhomogeneity coefficients estimated from FIDnavs were incorporated into an iterative reconstruction to retrospectively correct 2D gradient-echo images acquired in both axial and sagittal planes.
Results:
Spatiotemporal B0 field changes measured from rapidly acquired FIDnavs were in good agreement with the results of second-order SH fitting to the measured field maps. FIDnav field estimates accounted for a significant proportion of the ΔB0 variance induced by deep breathing (64 ± 21%) and nose touching (67 ± 34%) across all volunteers. Ghosting, blurring, and intensity modulation artifacts in -weighted images, induced by spatiotemporal field changes, were visibly reduced following retrospective correction with FIDnav inhomogeneity coefficients.
Conclusions:
Spatially resolved B0 inhomogeneity changes up to second order can be characterized in real time using the proposed approach. Retrospective FIDnav correction substantially improves -weighted image quality in the presence of strong B0 field modulations, with potential for real-time shimming.
Related Concept Videos
14:25Determining 3D Flow Fields via Multi-camera Light Field Imaging
07:50Electron Channeling Contrast Imaging for Rapid III-V Heteroepitaxial Characterization
08:18High-Accuracy Correction of 3D Chromatic Shifts in the Age of Super-Resolution Biological Imaging Using Chromagnon
06:53SpOT the Correct Tissue Every Time in Multi-tissue Blocks
10:04Sample Drift Correction Following 4D Confocal Time-lapse Imaging
11:57Measuring the Spin-Lattice Relaxation Magnetic Field Dependence of Hyperpolarized [1-13C]pyruvate