Evaluation of the Kejian KJ-8000 bilirubinometer in an Australian setting
Olivia Starowicz1, Philippa Edwards1, Peter Schmidt1
1Newborn Care Unit, Gold Coast University Hospital, Gold Coast, Queensland, Australia.
Aim:
To evaluate the accuracy of the Kejian 8000 (KJ-8000) transcutaneous bilirubinometer in infants of differing ethnicity and gestational age.
Methods:
This was a prospective study of infants in the Newborn Care Unit at Gold Coast University Hospital. Transcutaneous bilirubin (TcB) and serum bilirubin (SBR) results were compared using linear regression and a Bland-Altman plot. Predicted indices were calculated to assess the KJ-8000 as a screening tool using local jaundice management guidelines.
Results:
A total of 416 paired samples were collected from 201 infants. There was a strong correlation between TcB and SBR with a Pearson correlation coefficient of 0.8 (<0.00001). The bias was -5.9 μmol/L (95% confidence interval: -101, 89). The bias was not evenly spread, with the KJ-8000 tending to underestimate at higher SBR levels. Infants <32 weeks' gestation had a poor correlation of 0.48. Non-Caucasian infants were more likely to have TcB overestimation, and measurements were less precise. As a screening tool using local guidelines, the KJ-8000 had a sensitivity, specificity, positive predictive value and negative predictive value of 83, 53, 20 and 96%, respectively, and is predicted to avoid blood tests in 48% of infants screened.
Conclusions:
In this study, the correlation and agreement of TcB measurements using the KJ-8000 were not as good as has been reported with other more studied devices but may still have value as a screening tool. The poor correlation in preterm infants suggests that use should be restricted to term infants. The overall results of this study are affected by an underrepresentation of term infants, and so, further clinical assessment of this device should be undertaken before it can be recommended for widespread use.
Related Concept Videos
07:28Lipidico Injection Protocol for Serial Crystallography Measurements at the Australian Synchrotron
07:22Bacteriophage Effectiveness for Biocontrol of Foodborne Pathogens Evaluated via High-Throughput Settings
06:08Evaluation of Drug Sorption to PVC- and Non-PVC-based Tubes in Administration Sets Using a Pump
Setting Time of Cement
Documentation in Long-Term and Home Healthcare Setting
Long-Term Care Facilities
Specialized Care Centers and Settings-II
Rural health centers are specialized care facilities in remote locations with very few medical personnel. The primary care providers who run the centers are mostly Registered Nurse Practitioners. Here, emergency treatment is provided to critically ill or injured patients before they are transferred to the closest hospital. Fortunately, due to advancement in technology, many rural healthcare facilities and professionals have easy access to diagnostic and treatment...