Application of Vendor-Neutral Iterative Reconstruction Technique to Pediatric Abdominal Computed Tomography
Woo Hyeon Lim1,2, Young Hun Choi1,3, Ji Eun Park1
1Department of Radiology, Seoul National University Hospital, Seoul, Korea.
Objective:
To compare image qualities between vendor-neutral and vendor-specific hybrid iterative reconstruction (IR) techniques for abdominopelvic computed tomography (CT) in young patients.
Materials And Methods:
In phantom study, we used an anthropomorphic pediatric phantom, age-equivalent to 5-year-old, and reconstructed CT data using traditional filtered back projection (FBP), vendor-specific and vendor-neutral IR techniques (ClariCT; ClariPI) in various radiation doses. Noise, low-contrast detectability and subjective spatial resolution were compared between FBP, vendor-specific (i.e., iDose1 to 5; Philips Healthcare), and vendor-neutral (i.e., ClariCT1 to 5) IR techniques in phantom. In 43 patients (median, 14 years; age range 1-19 years), noise, contrast-to-noise ratio (CNR), and qualitative image quality scores of abdominopelvic CT were compared between FBP, iDose level 4 (iDose4), and ClariCT level 2 (ClariCT2), which showed most similar image quality to clinically used vendor-specific IR images (i.e., iDose4) in phantom study. Noise, CNR, and qualitative imaging scores were compared using one-way repeated measure analysis of variance.
Results:
In phantom study, ClariCT2 showed noise level similar to iDose4 (14.68-7.66 Hounsfield unit [HU] vs. 14.78-6.99 HU at CT dose index volume range of 0.8-3.8 mGy). Subjective low-contrast detectability and spatial resolution were similar between ClariCT2 and iDose4. In clinical study, ClariCT2 was equivalent to iDose4 for noise (14.26-17.33 vs. 16.01-18.90) and CNR (3.55-5.24 vs. 3.20-4.60) (p > 0.05). For qualitative imaging scores, the overall image quality ([reader 1, reader 2]; 2.74 vs. 2.07, 3.02 vs. 2.28) and noise (2.88 vs. 2.23, 2.93 vs. 2.33) of ClariCT2 were superior to those of FBP (p < 0.05), and not different from those of iDose4 (2.74 vs. 2.72, 3.02 vs. 2.98; 2.88 vs. 2.77, 2.93 vs. 2.86) (p > 0.05).
Conclusion:
Vendor-neutral IR technique shows image quality similar to that of clinically used vendor-specific hybrid IR technique for abdominopelvic CT in young patients.
More Related Videos
07:58Computational Reconstruction of Pancreatic Islets as a Tool for Structural and Functional Analysis
Published on: March 9, 2022
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
04:48Application of Deep Learning-Based Medical Image Segmentation via Orbital Computed Tomography
07:58Computational Reconstruction of Pancreatic Islets as a Tool for Structural and Functional Analysis
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
08:37Blood Circuit Reconstruction in an Abdominal Mouse Heart Transplantation Model
06:56Tree Core Analysis with X-ray Computed Tomography