Comparison of four inflation techniques on endotracheal tube cuff pressure using a feline airway simulator
Donna M White1, Mariano Makara1, Fernando Martinez-Taboada1
1University Veterinary Teaching Hospital, School of Veterinary Science, University of Sydney, Camperdown, Australia.
Objectives:
The aim of this study was to compare four inflation techniques on endotracheal tube cuff (ETC) pressure using a feline airway simulator.
Methods:
Ten participants used four different endotracheal cuff inflation techniques to inflate the cuff of a low-pressure, high-volume endotracheal tube within a feline airway simulator. The simulator replicated an average-sized feline trachea, intubated with a 4.5 mm endotracheal tube, connected to a circle breathing system and pressure-controlled ventilation with oxygen and medical air. Participants inflated the ETC: by pilot balloon palpation (P); by instilling the minimum occlusive volume (MOV) required for loss of airway leaks during mechanical ventilation; until a passive release of pressure with use of a loss-of-resistance syringe (LOR); and with use of a syringe with a digital pressure reader (D) specifically designed for endotracheal cuff inflation. Intracuff pressure was measured by a manometer obscured to participants. The ideal pressure was considered to be between 20 and 30 cmH2O. Data were analysed by Shapiro-Wilk, Kruskal-Wallis and χ2 tests, as appropriate.
Results:
Participants were eight veterinarians and two veterinary nurses with additional training in anaesthesia. Measured median intracuff pressures for P, MOV, LOR and D, respectively, were 25 cmH2O (range 4-74 cmH2O), 41 cmH2O (range 4-70 cmH2O), 31 cmH2O (range 18-64 cmH2O) and 22 cmH2O (range 20-30 cmH2O). D performed significantly better (P <0.001) than all other techniques, with no difference between the other techniques.
Conclusions And Relevance:
Use of D for cuff inflation achieved optimal cuff pressures. There may be high operator-dependent variability in the cuff pressures achieved with the use of P, MOV or LOR inflation techniques. As such, a cuff manometer is recommended when using any of these techniques.
Related Concept Videos
04:56Laryngeal Mask Airway (LMA) Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway (SGA)
COVID-19 / Coronavirus Outbreak: Protecting The Airway - Endotracheal Intubation
02:49Electromyographic Endotracheal Intubation in Pig: A Technique to Insert Electromyographic Tube Inside the Pig Trachea
04:30Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
02:58Tracheal Intubation for Mechanical Ventilation: A Procedure to Insert Endotracheal Tube in Trachea of Rabbit Model
08:26The Rigid Tube as an Alternative in Controlling the Problematic Airway