Validation of algorithmic CT image quality metrics with preferences of radiologists
Yuan Cheng1, Ehsan Abadi2,3, Taylor Brunton Smith1
1Clinical Imaging Physics Group, Medical Physics Graduate Program, Carl E. Ravin Advanced Imaging Laboratories, Duke University, 2424 Erwin Rd, Suite 302, Durham, NC, 27705, USA.
Purpose:
Automated assessment of perceptual image quality on clinical Computed Tomography (CT) data by computer algorithms has the potential to greatly facilitate data-driven monitoring and optimization of CT image acquisition protocols. The application of these techniques in clinical operation requires the knowledge of how the output of the computer algorithms corresponds to clinical expectations. This study addressed the need to validate algorithmic image quality measurements on clinical CT images with preferences of radiologists and determine the clinically acceptable range of algorithmic measurements for abdominal CT examinations.
Materials And Methods:
Algorithmic measurements of image quality metrics (organ HU, noise magnitude, and clarity) were performed on a clinical CT image dataset with supplemental measures of noise power spectrum from phantom images using techniques developed previously. The algorithmic measurements were compared to clinical expectations of image quality in an observer study with seven radiologists. Sets of CT liver images were selected from the dataset where images in the same set varied in terms of one metric at a time. These sets of images were shown via a web interface to one observer at a time. First, the observer rank ordered the CT images in a set according to his/her preference for the varying metric. The observer then selected his/her preferred acceptable range of the metric within the ranked images. The agreement between algorithmic and observer rankings of image quality were investigated and the clinically acceptable image quality in terms of algorithmic measurements were determined.
Results:
The overall rank-order agreements between algorithmic and observer assessments were 0.90, 0.98, and 1.00 for noise magnitude, liver parenchyma HU, and clarity, respectively. The results indicate a strong agreement between the algorithmic and observer assessments of image quality. Clinically acceptable thresholds (median) of algorithmic metric values were (17.8, 32.6) HU for noise magnitude, (92.1, 131.9) for liver parenchyma HU, and (0.47, 0.52) for clarity.
Conclusions:
The observer study results indicated that these algorithms can robustly assess the perceptual quality of clinical CT images in an automated fashion. Clinically acceptable ranges of algorithmic measurements were determined. The correspondence of these image quality assessment algorithms to clinical expectations paves the way toward establishing diagnostic reference levels in terms of clinically acceptable perceptual image quality and data-driven optimization of CT image acquisition protocols.
Related Concept Videos
08:04Monitoring Tumor Metastases and Osteolytic Lesions with Bioluminescence and Micro CT Imaging
07:05Area-based Image Analysis Algorithm for Quantification of Macrophage-fibroblast Cocultures
Micro-CT Imaging of a Mouse Spinal Cord
It's a little-known fact that the discovery and (inadvertent) use of X-rays garnered the first ever Nobel Prize in Physics. The famous X-ray image of Dr. Röntgen's wife's hand from 1895 that sent shock waves through the scientific community looks like most modern day 2D medical X-ray images. Though it is not the newest technology,...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
09:55Troubleshooting and Quality Assurance in Hyperpolarized Xenon Magnetic Resonance Imaging: Tools for High-Quality Image Acquisition
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice