In vivo measurement of pre-operational spallation source workers: baseline body burden levels and detection limits of relevant gamma emitters using high-resolution gamma spectrometry
Related Experiment Video
Updated: Jan 20, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Christopher L Rääf1, Anja Almén, Lena Johansson
1Medical Radiation Physics, Department of Translational Medicine (ITM), Lund University, SE-205 02, Malmö, Sweden.
Abstract:
As a measure to prepare for long-term internal dose monitoring of workers at the European Spallation Source (ESS) in Lund, Sweden, operated by the European Research Infrastructure Consortium (ERIC), as well as to enhance emergency preparedness against accidental releases, a series of in vivo measurements were conducted using a high-resolution HPGe detector with a 123% relative efficiency (1.332 MeV). This study describes the whole-body counting set-up, calibration procedure, and subsequent validation measurements using conventional NaI(Tl)-scanning-bed geometry on a selection of workers from the ESS. Detection limits for the relevant gamma emitters 7Be, 172Hf, and 182Ta were determined to be 65 Bq, 130 Bq, and 22 Bq, respectively, using a 2400 s acquisition time. The baseline measurements suggest that care must be taken to ensure that the fluctuations in the presence of radon daughters 214Bi and 214Pb are minimised by, for example, ensuring a minimum air exchange between the measuring room and the ambient air, and by demanding that the measured subjects change clothes and shower before measurement. Furthermore, in a monitoring program for internal doses to spallation source workers, the presence of radionuclides originating from non-work-related sources (such as 226Ra from private water wells or 137Cs from intakes of Chernobyl contaminated foodstuffs), or radionuclides from previous work history (such as 60Co within the nuclear power industry), must be considered.
More Related Videos
Related Concept Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
11:24Measuring DNA Damage and Repair in Mouse Splenocytes After Chronic In Vivo Exposure to Very Low Doses of Beta- and Gamma-Radiation
07:51Measurements of Soil Carbon by Neutron-Gamma Analysis in Static and Scanning Modes
09:25Detecting Pre-Stimulus Source-Level Effects on Object Perception with Magnetoencephalography
05:47Intravital Visualization of Gamma Delta Intraepithelial Lymphocytes
08:13Optimized Interferon-gamma ELISpot Assay to Measure T Cell Responses in the Guinea Pig Model after Vaccination