Rotational stability and refractive outcomes of a single-piece aspheric toric intraocular lens with 4 fenestrated
Giacomo Savini1, Giovanni Alessio2, Giuseppe Perone3
1GB Bietti Foundation IRCCS, Rome, Italy.
Purpose:
To assess the outcomes of implantation of a single-piece toric intraocular lens (IOL) with 4 fenestrated haptics.
Setting:
IRCCS Fondazione Bietti, Rome, Italy.
Design:
Prospective case series.
Methods:
All patients who had implantation of the Mini Toric Ready IOL were consecutively enrolled. Intraoperatively, the IOL was aligned using an automated system. Follow-up visits were performed at 1 day, 1 week, and 1, 3, and 6 months. At each visit, retroillumination pictures were taken to assess IOL orientation; visual acuity and refraction were also measured.
Results:
The final analysis comprised 63 eyes (63 patients). From the first to the last follow-up, the mean arithmetic rotation was -0.2 degrees ± 3.5 (SD) (range -13 to +10 degrees) and the mean absolute rotation was 1.6 ± 3.1 degrees. Intraocular lens rotation from the first to the last examination was within 5 degrees in 92.1% of eyes and on consecutive visits, within 5 degrees in 98.4% or more of eyes. By 6 months, 10 IOLs (15.9%) had rotated clockwise and 10 counterclockwise. Linear regression did not show a statistically significant relationship between rotational stability and the axis of placement with any preoperative parameter (eg, axial length). The mean magnitude of preexisting corneal astigmatism was 1.9 ± 0.7 diopters (D) (range 0.76 to 3.72 D). At the last follow-up, the mean magnitude of refractive astigmatism was 0.5 ± 0.4 D (range 0.0 to 1.5 D); the difference was statistically significant (P < .05).
Conclusion:
The toric IOL showed good rotational stability and is an option for correcting corneal astigmatism at the time of cataract surgery.
Related Concept Videos
04:59Rotating the Intraocular Lens to Prevent Posterior Capsular Opacification in Cataract Surgeries
11:49Synthesis of Soft Polysiloxane-urea Elastomers for Intraocular Lens Application
05:46Correction of Presbyopia by Monocular Bi-Aspheric Ablation Profile
Reflection and Refraction
Light travels at different speeds depending on the material through which it is propagating. When light travels from one material to another, it will either slow down or speed up. In order to conserve energy and momentum, the light must change the direction in which it propagates. This bending of light is known as refraction. Some fraction of...
08:16Adaptation of a Haptic Robot in a 3T fMRI
05:10Measuring the Behavioral Effects of Intraocular Scatter