Published on: July 19, 2013
Myocardial extracellular volume quantification in cardiac CT: comparison of the effects of two different iterative
Takafumi Emoto1, Masafumi Kidoh2, Seitaro Oda3
1Department of Central Radiology, Kumamoto University Hospital, 1-1-1 Honjyo, Chuo-ku, Kumamoto, 860-8556, Japan.
Objectives:
To compare the effects of hybrid iterative reconstruction (HIR) and model-based iterative reconstruction (MBIR) that incorporates a beam-hardening model for myocardial extracellular volume (ECV) quantification by cardiac CT using MRI as a reference standard.
Methods:
In this retrospective study, a total of 34 patients were evaluated using cardiac CT and MRI. Paired CT image sets were created using HIR and MBIR with a beam-hardening model. We calculated mean absolute differences and correlations between the global mid-ventricular ECV derived from CT and MRI via Pearson correlation analysis. In addition, we performed qualitative analysis of image noise and beam-hardening artifacts on postcontrast images using a four-point scale: 1 = extensive, 2 = strong, 3 = mild, and 4 = minimal.
Results:
The mean absolute difference between the ECV derived from CT and MRI for MBIR was significantly smaller than that for HIR (MBIR 3.74 ± 3.59%; HIR 4.95 ± 3.48%, p = 0.034). MBIR improved the correlation between the ECV derived from CT and MRI when compared with HIR (MBIR, r = 0.60, p < 0.001; HIR, r = 0.47, p = 0.006). In qualitative analysis, MBIR significantly reduced image noise and beam-hardening artifacts when compared with HIR ([image noise, MBIR 3.4 ± 0.7; HIR 2.1 ± 0.8, p < 0.001], [beam-hardening artifacts, MBIR 3.8 ± 0.4; HIR 2.6 ± 1.0, p < 0.001]).
Conclusions:
MBIR with a beam-hardening model effectively reduced image noise and beam-hardening artifacts and improved myocardial ECV quantification when compared with HIR using MRI as a reference standard.
Key Points:
• MBIR with a beam-hardening model effectively reduced image noise and beam-hardening artifacts. • The mean absolute difference between the global mid-ventricular ECV derived from CT and MRI for MBIR was significantly smaller than that for conventional HIR. • MBIR provided more accurate myocardial CT number and improved ECV quantification when compared with HIR.
Related Concept Videos
08:41Assessment of Cardiac Function and Myocardial Morphology Using Small Animal Look-locker Inversion Recovery (SALLI) MRI in Rats
10:46MRI and PET in Mouse Models of Myocardial Infarction
Combined SPECT and CT Imaging to Visualize Cardiac Functionality
Here we will demonstrate the fundamentals of single-photon emission computed tomography/computed tomography (SPECT/CT) imaging using mice. The technique involves injecting a radionuclide into a mouse, imaging the animal after it is distributed throughout the body, and then reconstructing the produced images to create a volumetric dataset. This can...
10:25Brain Infarct Segmentation and Registration on MRI or CT for Lesion-symptom Mapping
07:05Area-based Image Analysis Algorithm for Quantification of Macrophage-fibroblast Cocultures
12:03Myocardial Infarction and Functional Outcome Assessment in Pigs