Optical imaging provides rapid verification of static small beams, radiosurgery, and VMAT plans with millimeter
Muhammad Ramish Ashraf1, Petr Bruza1, Brian W Pogue1,2
1Thayer School of Engineering, Dartmouth College Hanover, Hanover, NH, 03755, USA.
Purpose:
We demonstrate the feasibility of optical imaging as a quality assurance tool for static small beamlets, and pretreatment verification tool for radiosurgery and volumetric-modulated arc therapy (VMAT) plans.
Methods:
Small static beams and clinical VMAT plans were simulated in a treatment planning system (TPS) and delivered to a cylindrical tank filled with water-based liquid scintillator. Emission was imaged using a blue-sensitive, intensified CMOS camera time-gated to the linac pulses. For static beams, percentage depth and cross beam profiles of projected intensity distribution were compared to TPS data. Two-dimensional (2D) gamma analysis was performed on all clinical plans, and the technique was tested for sensitivity against common errors (multileaf collimator position, gantry angle) by inducing deliberate errors in the VMAT plans control points. The technique's detection limits for spatial resolution and the smallest number of control points that could be imaged reliably were also tested. The sensitivity to common delivery errors was also compared against a commercial 2.5D diode array dosimeter.
Results:
A spatial resolution of 1 mm was achieved with our imaging setup. The optical projected percentage depth intensity profiles agreed to within 2% relative to the TPS data for small static square beams (5, 10, and 50 mm2 ). For projected cross beam profiles, a gamma pass rate >99% was achieved for a 3%/1 mm criteria. All clinical plans passed the 3%/3 mm criteria with >95% passing rate. A static 5 mm beam with 20 Monitor Units could be measured with an average percent difference of 5.5 ± 3% relative to the TPS. The technique was sensitive to multileaf collimator errors down to 1 mm and gantry angle errors of 1°.
Conclusions:
Optical imaging provides ample spatial resolution for imaging small beams. The ability to faithfully image down to 20 MU of 5 mm, 6 MV beamlets prove the ability to perform quality assurance for each control point within dynamic plans. The technique is sensitive to small offset errors in gantry angles and multileaf collimator (MLC) leaf positions, and at certain scenario, it exhibits higher sensitivity than a commercial 2.5D diode array.
Related Concept Videos
08:31Three-dimensional Optical-resolution Photoacoustic Microscopy
13:49High-resolution Fiber-optic Microendoscopy for in situ Cellular Imaging
10:17Switchable Acoustic and Optical Resolution Photoacoustic Microscopy for In Vivo Small-animal Blood Vasculature Imaging
04:35A Rapid Approach to High-Resolution Fluorescence Imaging in Semi-Thick Brain Slices
08:49Rapid Whole-Mount High-Resolution Imaging of Small Animal Vasculature for Quantitative Studies
08:25Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System