Novel methods for the quantification of toxic, residual phase transfer catalyst in fluorine-18 labeled radiotracers
David W Blevins1, Grant H Rigney2, Michael Y Fang2
1The University of Tennessee Graduate School of Medicine, 1924 Alcoa Hwy., Knoxville, TN 37920, United States of America.
Introduction:
Fluorine-18 labeled radiopharmaceuticals undergo quality control testing for residual phase-transfer-catalyst content. The almost universally used quality-control test is a silica plate spot-test comparison of the radiopharmaceutical beside a 50-ppm standard. Once developed by staining, the radiopharmaceutical spot must be of equal or less intensity to pass the test. There is currently a need for a quantitative, inexpensive, and less subjective quality control method that allows the automatic incorporation of the acquired measurement directly into electronic batch reports.
Results:
In the developed method, a resazurin test solution is mixed with an aliquot of the radiopharmaceutical analyte along with dichloromethane (DCM). The mixture is vortexed. The potassium resazurin-phase transfer catalyst complex solubilizes into the DCM imparting a blue color. The organic layer is then removed for analysis. Three measurement methods were utilized: visual colorimetry against pre-prepared standards, spectrophotometric measurement of transmittance, and electrical conductance. A simple prototype spectrophotometer and an electrical test cell were constructed to acquire data. Sodium Resazurin dye was found to be a suitable test chromophore for residual phase transfer catalyst analysis of aqueous solutions. Quantitative spectrophotometric measurements are possible in the 0-100-ppm range (18-crown-6) and 0-150-ppm range (Kryptofix® or tetrabutylammonium). Electrical resistance measurements of the phase transfer-catalyst resazurin complex in DCM are also a viable method, allowing quantitative phase transfer catalyst measurements in the 0-100-ppm range.
Conclusion:
The methodologies developed are more quantitative alternatives to the current spot-test method. The spectrophotometric method was determined to be the most accurate method.
More Related Videos
Related Concept Videos
09:5718F-Labeling of Radiotracers Functionalized with a Silicon Fluoride Acceptor (SiFA) for Positron Emission Tomography
10:20Automation of a Positron-emission Tomography (PET) Radiotracer Synthesis Protocol for Clinical Production
11:03Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases
10:56Metabolic Labeling and Profiling of Transfer RNAs Using Macroarrays
08:33Radiosynthesis of 1-(2-[18F]Fluoroethyl)-L-Tryptophan using a One-pot, Two-step Protocol
09:53Predicting Catalyst Extrudate Breakage Based on the Modulus of Rupture

![Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55537.jpg&w=3840&q=50)