The Effect of Piezoelectric Ultrasonic Scaler Tip Wear on Root Surface Roughness at Different Working Parameters: An
Hariharasudhan Vengatachalapathi1, Rajesh Naik2, Rachana Rao2
1Department of Periodontology, AECS Maaruti College of Dental Sciences and Research Centre, Bangalore, India. dr.hariharasudhan@gmail.com.
Objective:
This study aimed to evaluate the influence of scaler tip wear and different working parameters, i.e., lateral force, power setting and tip angulation, on the roughness of root surfaces following treatment with piezoelectric ultrasonic scaling devices.
Materials:
Twenty piezoelectric ultrasonic scaler inserts (10 worn/10 new) were selected to examine the erosion ratio (ER) under atomic force microscopy (AFM). A total of 160 root samples were prepared and instrumented by new (n = 80) and worn inserts (n = 80) at different working parameters (tip angulation, power setting, lateral force). Roughness change (Rc) on root surfaces after instrumentation was examined by a contact profilometer.
Results:
Statistically significant differences were found between the mean ERs of new and worn tips (p < 0.0001). The various combinations of the assessed working parameters showed synergistic effects resulting in a wide range of root surface roughness. The present study found the higher Rc in the group with a 45° angulation, (P10) high power setting and 1.0 N lateral force (subgroup 8) when compared to other groups. Among the groups, the worn scaler tips subgroup 8 showed a higher Rc (5.692 ± 0.81) when compared to new scaler tips subgroup 8 (4.798 ± 0.51; p < 0.001).
Conclusion:
The findings of the present study highlighted that scaler tip wear strongly influences the root surface roughness when used at higher tip angulation, lateral force and power settings. Hence, ultrasonic scaler tip wear should be periodically evaluated and should be considered as much as the other working parameters.
Related Concept Videos
05:30Roughness Impact of Piezoelectric Dental Scaler on Two Distinct Flowable Composite Filling Materials
09:21Experiments on Ultrasonic Lubrication Using a Piezoelectrically-assisted Tribometer and Optical Profilometer
06:45Force Spectroscopy of Single Protein Molecules Using an Atomic Force Microscope
05:51Atomic Force Microscopy to Study the Physical Properties of Epidermal Cells of Live Arabidopsis Roots
08:31Probing Surface Electrochemical Activity of Nanomaterials using a Hybrid Atomic Force Microscope-Scanning Electrochemical Microscope (AFM-SECM)
09:27Automation of Bio-Atomic Force Microscope Measurements on Hundreds of C. albicans Cells