Biomechanical Comparison of Fixed- versus Variable-Angle Locking Screws for Distal Humerus Comminuted Fractures
Ali Nourbakhsh1, Adam G Hirschfeld2, Sravan Dhulipala1
1Department of Orthopedics, WellStar Atlanta Medical Center, Atlanta, GA, USA.
Background:
To compare the stability of fixed- versus variable-angle locking constructs for the comminuted distal humerus fracture (AO/OTA 13-A3).
Methods:
Eight pairs of complete humeri harvested from eight fresh frozen cadavers were used for the study. We fixed the intact humeri using 2.7-mm/3.5-mm locking VA-LCP stainless steel distal humerus posterolateral (nine-hole) and medial (seven-hole) plates. An oscillating saw was used to cut a 1-cm gap above the olecranon fossa. The specimens were loaded in axial mode with the rate of 1 mm per 10 seconds to failure, and stress-strain curves were compared in each pair. The mode of failure was recorded as well as the load needed for 2- and 4-mm displacement at the lateral end of the gap.
Results:
The stiffness of the constructs, based on the slope of the stress-strain curve, did not show any difference between the fixed- versus variable-angle constructs. Likewise, there was no difference between the force needed for 2- or 4-mm displacement at the lateral gap between the fixed- and variable-angle constructs. The mode of failure was bending of both plates in all specimens and screw pull-out in four specimen pairs in addition to the plate bending.
Conclusions:
Our results did not show any difference in the biomechanical stability of the fixed- versus variable-angle constructs. There was not any screw breakage or failure of the plate-screw interface.
Related Concept Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
10:09Biomechanical Testing of Murine Tendons