Adaptive attenuation correction during H-scan ultrasound imaging using K-means clustering
Haowei Tai1, Mawia Khairalseed2, Kenneth Hoyt3
1Department of Electrical and Computer Engineering, University of Texas at Dallas, Richardson, TX, USA; Department of Biomedical Engineering, University of Texas at Dallas, Richardson, TX, USA.
Abstract:
H-scan ultrasound (US) imaging (where the 'H' stands for Hermite) is a novel non-invasive, low cost and real-time technology. Like traditional US, H-scan US suffers from frequency-dependent attenuation that must be corrected to have acceptable image quality for tissue characterization. The goal of this research was to develop a novel attenuation correction method based on adaptive K-means clustering. To properly isolate these signals, a lateral moving window approach applied to adaptively adjust GH filters based on the changing of RF vector spectrums. Then the signal isolated via the same filter will be combined together via overlap-add technology to keep the information loss minimum. Experimental data was collected using a Verasonics 256 US scanner equipped with a L11-4v linear array transducer. In vivo data indicates that H-scan US imaging after adaptive attenuation correction can optimally re-scale the GH kernels and match to the changing spectrum undergoing attenuation (i.e. high frequency shift). This approach produces H-scan US images with more uniform spatial intensity and outperforms global attenuation correction strategies. Overall, this approach will improve the ability of H-scan US imaging to estimate acoustic scatterer size and will improve its clinical use for tissue characterization when imaging complex tissues.
Related Concept Videos
07:34Delivery of Antibodies into the Brain Using Focused Scanning Ultrasound
10:02High-frequency Ultrasound Imaging of Mouse Cervical Lymph Nodes
High-frequency Ultrasound Imaging of the Abdominal Aorta
High-frequency ultrasound systems are used to acquire high resolution images. Here, the use of a state-of-the-art system will be demonstrated to image the morphology and hemodynamics of small pulsatile arteries and veins found in mice and rats. Ultrasound is a relatively inexpensive, portable, and versatile method for the noninvasive...
10:21A Coregistered Ultrasound and Photoacoustic Imaging Protocol for the Transvaginal Imaging of Ovarian Lesions
10:53A Novel Application of Musculoskeletal Ultrasound Imaging
08:21Wideband Optical Detector of Ultrasound for Medical Imaging Applications