Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Anchored Transponder Guided Lung Radiation Therapy
Michael C Dobelbower1, Richard A Popple1, Douglas J Minnich2
1Department of Radiation Oncology, University of Alabama at Birmingham, Birmingham, Alabama.
Purpose:
The Calypso Beacon transponder has been modified by the addition of a nitinol anchor feature to allow for positional stability when implanted bronchoscopically into the lung. The purpose of this study was to confirm the feasibility and safety of anchored transponder placement and feasibility of lung target localization and tracking.
Methods And Materials:
This study enrolled patients with histologically confirmed cancer in the lung (primary or metastatic) who were scheduled to receive external beam radiation therapy. Three anchored transponders were implanted via flexible bronchoscopy into small (approximately 2- to 2.5-mm diameter) airways. Patient alignment at each radiation fraction was performed with the Calypso system, and anchored transponder position was tracked during radiation delivery. The primary endpoint was defined as the ability to localize at least 85% of the patients during the first week of treatment. Four follow-up visits were specified including a posttreatment assessment and every 3 months up to 1 year.
Results:
A total of 69 patients underwent anchored transponder placement, and all 207 implanted anchored transponders were visible on the treatment-planning simulation computed tomography scan. Sixty-seven patients underwent radiation therapy, and localization was successful in 66 cases (98.5%). With 1 failure in 67 cases, the P value for rejecting the null hypothesis was <.001 and the primary objective of the study met. Five adverse events in 5 patients were potentially attributed to the study device or implantation procedure, consisting of pneumonia (2 cases), pleural abscess (1 case), and pneumothorax (2 cases). Two serious events (cardiac arrest and acute hypotension) were attributed to anesthesia during the implantation procedure.
Conclusions:
This study strongly supports that anchored transponders are safe, positionally stable, and useful for lung tumor localization and monitoring.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
04:04Endobronchial Ultrasound-guided Intratumoral Injection of Cisplatin for the Treatment of Isolated Mediastinal Recurrence of Lung Cancer
Published on: February 12, 2017
Related Concept Videos
08:17Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
The Anchoring-and-Adjustment Heuristic
04:04Endobronchial Ultrasound-guided Intratumoral Injection of Cisplatin for the Treatment of Isolated Mediastinal Recurrence of Lung Cancer
08:03Anticancer Efficacy of Photodynamic Therapy with Lung Cancer-Targeted Nanoparticles
05:56Phase-Resolved Functional Lung MRI for Pulmonary Ventilation and Perfusion (V/Q) Assessment