Additional value of triple-sensor urethral catheter in demonstrating urethral pressure variations during filling
M T Kummeling1, E Bovelander2, J I van Uhm3
1Department of Urology, Haaglanden MC, The Hague, The Netherlands.
Aims:
During filling cystometry, urethral pressure variations (UPV) can be observed. The clinical relevance and a clear definition of this phenomenon are still a matter of debate. For further research and definition of UPV, it is important to determine how this condition can best be demonstrated. The purpose of this study is to compare continuous urethral pressure measurements with a single urethral-sensor catheter and a triple urethral-sensor catheter in demonstrating UPV.
Methods:
Seventy-five adult female patients requiring urodynamic investigation enrolled in this prospective study. All patients underwent two series of filling and voiding cystometry. One series was performed with a dual-air balloon sensor urodynamic catheter, the other series with a triple urethral-sensor catheter. UPV were defined as urethral pressure drop exceeding 30 cmH2O.
Results:
The prevalence of UPV was 37.3% (28 out of 75 patients), more common than detrusor overactivity. The triple urethral-sensor catheter was more sensitive than the single urethral-sensor catheter: In eight patients UPV were demonstrated with both catheters and in 18 patients only in the measurement with the triple urethral-sensor catheter. This difference in detection was significant (P < .001).
Conclusion:
There is an additional value in measurement with the triple urethral-sensor catheter for demonstration of UPV during filling cystometry. Currently, continuous measurement of urethral pressure during filling cystometry and UPV is not defined within International Continence Society terminology. The single urethral-sensor catheter is useful for a start, however, it demonstrates less than half of all UPV.
Related Concept Videos
10:07Cystometric and External Urethral Sphincter Measurements in Awake Rats with Implanted Catheter and Electrodes Allowing for Repeated Measurements
05:09Urethral Stricture Induction Followed by Buccal Mucosa Graft Urethroplasty in a Rat Model
07:42Nerve-sparing Mid-urethral Obstruction (NeMO) in Female Small Rodents
09:31Evaluating the Procedure for Performing Awake Cystometry in a Mouse Model
03:55Urethroplasty with Pedicled Tunica Vaginalis for the Treatment of Long-segment Anterior Urethral Stricture Caused by Lichen Sclerosus of Glans Penis
Variation of Atmospheric Pressure
Assuming the air temperature is constant at a given altitude and that the ideal gas law of thermodynamics describes the atmosphere to a good approximation, one can find the variation of atmospheric pressure with height.
Let p(y) be the atmospheric pressure at...