A 12-year single-center retrospective analysis of antisiphon devices to prevent proximal ventricular shunt
Joyce Koueik1, Mark R Kraemer1, David Hsu2
1Departments of1Neurological Surgery.
Objective:
Recent evidence points to gravity-dependent chronic shunt overdrainage as a significant, if not leading, cause of proximal shunt failure. Yet, shunt overdrainage or siphoning persists despite innovations in valve technology. The authors examined the effectiveness of adding resistance to flow in shunt systems via antisiphon devices (ASDs) in preventing proximal shunt obstruction.
Methods:
A retrospective observational cohort study was completed on patients who had an ASD (or additional valve) added to their shunt system between 2004 and 2016. Detailed clinical, radiographic, and surgical findings were examined. Shunt failure rates were compared before and after ASD addition.
Results:
Seventy-eight patients with shunted hydrocephalus were treated with placement of an ASD several centimeters distal to the primary valve. The records of 12 of these patients were analyzed separately due to a complex shunt revision history (i.e., > 10 lifetime shunt revisions). The authors found that adding an ASD decreased the 1-year ventricular catheter obstruction rates in the "simple" and "complex" groups by 67.3% and 75.8%, respectively, and the 5-year rates by 43.3% and 65.6%, respectively. The main long-term ASD complication was ASD removal for presumed valve pressure intolerance in 5 patients.
Conclusions:
Using an ASD may result in significant reductions in ventricular catheter shunt obstruction rates. If confirmed with prospective studies, this observation would lend further evidence that chronic shunt overdrainage is a central cause of shunt malfunction, and provide pilot data to establish clinical and laboratory studies that assess optimal ASD type, number, and position, and eventually develop shunt valve systems that are altogether resistant to siphoning.
Related Concept Videos
14:59Neuronavigation and Laparoscopy Guided Ventriculoperitoneal Shunt Insertion for the Treatment of Hydrocephalus
09:15In Vitro Thrombosis Test for Ventricular Assist Devices
08:49Use of Two Intracorporeal Ventricular Assist Devices As a Total Artificial Heart
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
09:23Shunt Surgery, Right Heart Catheterization, and Vascular Morphometry in a Rat Model for Flow-induced Pulmonary Arterial Hypertension
05:06Mechanical Dissociation of Tissues for Single Cell Analysis Using a Motorized Device