Related Experiment Video
Updated: Jan 19, 2026
Cross-Sectional Research: Parallel Study of Multiple Cohorts
Reliability of a Smartphone Compared With an Inertial Sensor to Measure Shoulder Mobility: Cross-Sectional Study
Cristina Roldán-Jiménez1,2, Jaime Martin-Martin1,3, Antonio I Cuesta-Vargas1,2,4
1Clinimetric Group F-14 Biomedical Research Institute of Malaga, Malaga, Spain.
Background:
The shoulder is one of the joints with the greatest mobility within the human body and its evaluation is complex. An assessment can be conducted using questionnaires or functional tests, and goniometry can complement the information obtained in this assessment. However, there are now validated devices that can provide more information on the realization of movement, such as inertial sensors. The cost of these devices is usually high and they are not available to all clinicians, but there are also inertial sensors that are implemented in mobile phones which are cheaper and widely available. Results from the inertial sensors integrated into mobile devices can have the same reliability as those from dedicated sensors.
Objective:
This study aimed to validate the use of the Nexus 4 smartphone as a measuring tool for the mobility of the humerus during shoulder movement compared with a dedicated InertiaCube3 (Intersense) sensor.
Methods:
A total of 43 subjects, 27 affected by shoulder pathologies and 16 asymptomatic, participated in the study. Shoulder flexion, abduction, and scaption were measured using an InertiaCube3 and a Nexus 4 smartphone, which were attached to the participants to record the results simultaneously. The interclass correlation coefficient (ICC) was calculated based on the 3 movements performed.
Results:
The smartphone reliably recorded the velocity values and simultaneously recorded them alongside the inertial sensor. The ICCs of the 3 gestures and for each of the axes of movement were analyzed with a 95% CI. In the abduction movement, the devices demonstrated excellent interclass reliability for the abduction humeral movement axis (Cronbach alpha=.98). The axis of abduction of the humeral showed excellent reliability for the movements of flexion (Cronbach alpha=.93) and scaption (Cronbach alpha=.98).
Conclusions:
Compared with the InertiaCube3, the Nexus 4 smartphone is a reliable and valid tool for recording the velocity produced in the shoulder.
Related Concept Videos
Cross-Sectional Research
10:10Reverse Total Shoulder Arthroplasty
07:00Measuring the Switch Cost of Smartphone Use While Walking
Shoulder Exam I
Examination of the shoulder can be complex, because it actually consists of four separate joints: are the glenohumeral (GH) joint, the acromioclavicular (AC) joint, the sternoclavicular joint, and the scapulothoracic joint. The GH joint is primarily responsible for shoulder motion and is the most mobile joint in the body. It has been likened to a golf ball sitting on a tee and is prone to instability. It is held in place...
07:51Video Movement Analysis Using Smartphones (ViMAS): A Pilot Study
16:11Thermochemical Studies of Ni(II) and Zn(II) Ternary Complexes Using Ion Mobility-Mass Spectrometry