Related Experiment Video
Updated: Jan 19, 2026

Gene-therapy Inspired Polycation Coating for Protection of DNA Origami Nanostructures
Published on: January 19, 2019
Delivery of targeted gene therapies using a hybrid cryogel-coated prosthetic vascular graft
Cindy Huynh1,2, Ting-Yu Shih3,4, Alexander Mammoo1,3,5
1Division of Vascular and Endovascular Surgery, Beth Israel Deaconess Medical Center, Boston, MA, United States of America.
Objectives:
The success of prosthetic vascular grafts in the management of peripheral arterial disease is frequently limited by the development of anastomotic neointimal hyperplasia (ANIH), with the host response to prosthetic grafts beginning soon after implantation. To address this, we combine a platform of polyethylene terephthalate (PET) fabric with an applied cryogel layer containing biologic agents to create a bioactive prosthetic graft system, with the ability to deliver therapeutics targeting modulators of the ANIH-associated transcriptome response, along with antithrombotic agents.
Methods:
Hybrid graft materials were synthesized by cryopolymerization of methacrylated alginate and heparin onto electrospun (ePET), knitted PET (kPET), or woven PET (wPET). Arg-Gly-Asp (RGD) peptides were added to increase cell adhesion. Scanning electron microscopy (SEM) was used to study the microstructure at 1 day, and 2, 4, and 8 weeks. Physical properties such as swelling ratio, pore connectivity, shape recovery, and stiffness were evaluated. Human aortic endothelial cell (HAoEC) adherence was visualized using confocal microscopy after 24 hours and proliferation was evaluated with a resazurin-based assay for 7 days. Confocal microscopy was used to assess delivery of adeno-associated virus (AAV-GFP) after incubation of hybrid grafts with HAoECs. Heparin activity of the materials was measured using an anti-Xa assay.
Results:
SEM demonstrated large interconnected pores throughout the entire structure for all graft types, with minimal degradation of the cryogel after 8 weeks. Hybrid materials showed a trend towards increased shape recovery, increased stiffness, decreased swelling ratio, and no difference in pore connectivity. HAoECs incorporated, adhered, and proliferated over 7 days on all materials. HAoECs were successfully transduced with AAV-GFP from the hybrid graft materials. Anti-Xa assay confirmed continued activity of heparin from all materials for over 7 days.
Conclusions:
We have developed a bioactive prosthetic graft system with a cryogel coating capable of delivering biologic agents with antithrombotic activity. By applying the cryogel and selected agents onto PET prior to graft implantation, this study sets the stage for the system to be individualized and tailored to the patient, with bioengineering and targeted gene therapy strategies dovetailing to create an improved prosthetic graft adaptable to emerging knowledge and technologies.
Related Concept Videos
08:30Gene-therapy Inspired Polycation Coating for Protection of DNA Origami Nanostructures
Gene Therapy
08:04Customizing a Cryolite Glass Prosthetic Eye
05:09Lumbar Intrathecal Injection of Gene Therapy Vectors for Central Nervous System Targeting in Mice and Rats
11:22Engineering Biological-Based Vascular Grafts Using a Pulsatile Bioreactor
10:38Synthesis of Functionalized 10-nm Polymer-coated Gold Particles for Endothelium Targeting and Drug Delivery