Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope10:25

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope

10.6K
We use an aberration-corrected scanning transmission electron microscope to define single-digit nanometer patterns in two widely-used electron-beam resists: poly (methyl methacrylate) and hydrogen silsesquioxane. Resist patterns can be replicated in target materials of choice with single-digit nanometer fidelity using liftoff, plasma etching, and resist infiltration by...
10.6K
Correction of Presbyopia by Monocular Bi-Aspheric Ablation Profile05:46

Correction of Presbyopia by Monocular Bi-Aspheric Ablation Profile

781
This article provides a step-by-step guide to correcting presbyopia with a monocular bi-aspheric ablation...
781
Spherical Coordinates01:23

Spherical Coordinates

14.8K
Spherical coordinate systems are preferred over Cartesian, polar, or cylindrical coordinates for systems with spherical symmetry. For example, to describe the surface of a sphere, Cartesian coordinates require all three coordinates. On the other hand, the spherical coordinate system requires only one parameter: the sphere's radius. As a result, the complicated mathematical calculations become simple. Spherical coordinates are used in science and engineering applications like electric and...
14.8K
Exploring X Chromosomal Aberrations in Ovarian Cells by Using Fluorescence In Situ Hybridization11:08

Exploring X Chromosomal Aberrations in Ovarian Cells by Using Fluorescence In Situ Hybridization

1.4K
This article presents two methods based on fluorescence in situ hybridization to determine the X chromosomal content of ovarian cells in non-grafted and grafted ovarian cortex tissue from females with X chromosomal...
1.4K
Distance Corrections01:15

Distance Corrections

281
To achieve precise distance measurements, especially in surveying and construction, certain corrections must be applied to account for potential sources of error like the standardization errors, temperature variations, and slope adjustments.Standardization error emerges when measurement equipment undergoes changes, such as wear, repairs, or weather impacts. To address this, surveyors compare the equipment’s readings to a standard. This process identifies any deviation that might lead to...
281
Power Factor Correction01:20

Power Factor Correction

491
The power transmission to a factory involves the transfer of apparent power, a combination of active and reactive power. The power factor measures how effectively electrical power is converted into useful work output. The ratio of the real power (KW) that does the work to the apparent power (KVA) supplied to the circuit.
491

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Simplification of a Three-Constant Intraocular Lens Calculation Formula to a Single-Constant Approach: The Haigis Formula.

Diagnostics (Basel, Switzerland)·2026
Same author

Machine Learning-Based Prediction of Long-Term Intraocular Pressure Fluctuations in Open-Angle Glaucoma.

Ophthalmology science·2026
Same author

Intraocular Lens Calculation Concept Based on Aphakic Refraction-Considerations on a Cornea Model With Two Refracting Surfaces.

Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists)·2026
Same author

Refractive and Corneal Astigmatism After Implantation of a Supraciliary Drainage Device in Microinvasive Glaucoma Surgery.

Journal of refractive surgery (Thorofare, N.J. : 1995)·2026
Same author

Correction: Prediction of the ectasia screening index from raw Casia2 volume data for keratoconus identification by using convolutional neural networks.

PloS one·2026
Same author

Response to the Letter to the Editor: Dual-Zone Keratometry for Identifying Central Radius and Corneal Asphericity.

Current eye research·2026

Related Experiment Video

Updated: Jan 19, 2026

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope
10:25

Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope

Published on: September 14, 2018

10.6K

Individually Customized IOL Versus Standard Spherical Aberration-Correcting IOL.

Jens Schrecker, Simon Schröder, Achim Langenbucher

    Journal of Refractive Surgery (Thorofare, N.J. : 1995)
    |September 10, 2019
    PubMed
    Summary

    This study compared a new, personalized lens for cataract surgery against a standard lens. Patients receiving the custom-made lens showed better vision at various distances and improved contrast sensitivity compared to those with the standard lens. These findings suggest that tailoring lens design to individual eye characteristics may enhance visual quality after surgery.

    Keywords:
    Intraocular LensSpherical AberrationContrast SensitivityVisual Acuity

    Frequently Asked Questions

    More Related Videos

    Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
    05:46

    Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity

    Published on: September 20, 2024

    781
    Exploring X Chromosomal Aberrations in Ovarian Cells by Using Fluorescence In Situ Hybridization
    11:08

    Exploring X Chromosomal Aberrations in Ovarian Cells by Using Fluorescence In Situ Hybridization

    Published on: April 7, 2023

    1.4K

    Related Experiment Videos

    Last Updated: Jan 19, 2026

    Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope
    10:25

    Single-Digit Nanometer Electron-Beam Lithography with an Aberration-Corrected Scanning Transmission Electron Microscope

    Published on: September 14, 2018

    10.6K
    Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
    05:46

    Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity

    Published on: September 20, 2024

    781
    Exploring X Chromosomal Aberrations in Ovarian Cells by Using Fluorescence In Situ Hybridization
    11:08

    Exploring X Chromosomal Aberrations in Ovarian Cells by Using Fluorescence In Situ Hybridization

    Published on: April 7, 2023

    1.4K

    Area of Science:

    • Ophthalmology and visual science research within Individually Customized IOL technology
    • Refractive surgery outcomes research within clinical optics

    Background:

    Cataract surgery requires precise lens selection to restore clear vision for patients. Standard lenses often use a fixed correction for spherical aberration to improve image quality. However, individual eyes vary significantly in their corneal shape and optical properties. No prior work had fully resolved whether personalized lens designs outperform these conventional fixed-correction models. That uncertainty drove the need for a direct clinical comparison between these two approaches. Prior research has shown that residual optical errors can limit visual performance after standard procedures. This gap motivated the current investigation into custom-designed implants. The study addresses how specific patient measurements influence the final visual outcome.

    Purpose Of The Study:

    This study aimed to compare the visual performance of an individually customized intraocular lens against a standard spherical aberration-correcting model. Surgeons often seek ways to improve patient outcomes beyond traditional, non-personalized lens selection methods. The researchers investigated whether incorporating patient-specific data into lens calculations yields better results. This work addresses the limitations inherent in using fixed corrections for all patients regardless of their unique ocular anatomy. The team sought to determine if a merit function approach could effectively optimize refractive outcomes. They also aimed to quantify differences in contrast sensitivity and defocus curves between the two groups. By comparing these distinct methodologies, the authors hoped to clarify the benefits of personalized optical design. This investigation provides evidence regarding the efficacy of custom-made implants in modern clinical practice.

    Main Methods:

    The investigators conducted a prospective comparative trial involving 74 eyes from 60 patients. Participants underwent randomization in a 2:1 ratio to receive either the novel custom lens or a standard aspheric implant. Clinical staff performed comprehensive assessments at four weeks and three months after the operation. The review approach included measuring refraction and visual acuity at far, intermediate, and near distances. Technical teams evaluated contrast sensitivity under both photopic and mesopic lighting environments. Researchers also mapped the defocus curve to determine the range of clear vision. Corneal and ocular spherical aberration measurements provided data on optical quality. Finally, the team recorded pupil dimensions to ensure accurate comparison of the two lens groups.

    Main Results:

    The customized group demonstrated significantly better uncorrected distance visual acuity and distance-corrected near visual acuity at the three-month mark. Contrast sensitivity testing revealed superior results for the custom group across nearly all spatial frequencies. Under both photopic and mesopic lighting, the personalized lens outperformed the standard model. The defocus curve for the custom group displayed a wider plateau surrounding the distance focal point. Mean total ocular spherical aberration at 5 mm was 0.04 ± 0.06 µm for the custom group. The standardized group showed a mean value of -0.01 ± 0.05 µm for the same metric. These findings indicate that personalized lens design provides a measurable improvement in visual quality. The data confirm that individual optimization leads to more favorable optical outcomes than fixed corrections.

    Conclusions:

    The authors suggest that personalized implants offer superior visual performance over standard options. Their data indicate that customizing lens parameters leads to better distance and near vision. The findings imply that contrast sensitivity benefits significantly from this tailored approach. The researchers propose that the wider plateau in the defocus curve represents a functional advantage for patients. Their analysis shows that individual optimization effectively reduces total ocular spherical aberration. The study supports the use of custom calculations to refine surgical results. These results provide evidence that patient-specific modeling improves overall optical quality. The authors conclude that individual lens design is a viable strategy for enhancing post-surgical outcomes.

    The customized group achieved significantly better uncorrected distance and distance-corrected near visual acuity. Furthermore, these patients demonstrated superior contrast sensitivity across most spatial frequencies under both photopic and mesopic conditions compared to the standardized group.

    The customized lens calculation utilized a merit function incorporating residual refraction, residual spherical aberration, and modulation transfer function. In contrast, the standardized group relied on the Holladay formula with a fixed spherical aberration correction of -0.27 µm.

    The researchers state that a 5 mm pupil size was necessary to evaluate the mean total ocular spherical aberration. This specific aperture allowed for consistent measurement of optical performance across all participants at the three-month follow-up.

    The study utilized a prospective comparative design to analyze 74 eyes. Researchers randomized participants in a 2:1 ratio, with 48 eyes receiving the custom lens and 26 eyes receiving the standard lens to assess performance differences.

    The customized group exhibited a mean total ocular spherical aberration of 0.04 ± 0.06 µm, whereas the standardized group measured -0.01 ± 0.05 µm. This measurement highlights the effectiveness of the personalized approach in managing ocular aberrations.

    The authors propose that implanting an individually optimized aspheric lens significantly improves visual performance. They specifically highlight enhanced contrast sensitivity as a primary benefit of this personalized surgical strategy over standard aberration-correcting models.