Application of Low-dose Stereoradiography in In Vivo Vertebral Morphologic Measurements: Comparison With Computed
Saba Pasha1, Tom Schlösser2, Xiaowei Zhu3
1Division of Orthopaedic Surgery.
Background:
Though computed tomography (CT) and 3 dimensional (3D) reconstruction of the spine and ribcage are powerful techniques for detailed monitoring of spinal growth and surgical planning of patients, drawbacks can arise. We explored the application of low-dose stereoradiography of the spine to calculate distinct morphologic parameters of the vertebral body in a juvenile patient population with early-onset scoliosis or congenital scoliosis. This study compares the 3D vertebral morphology measurements using low-dose stereoradiography with the currently accepted imaging modality for such measurements, CT scans.
Methods:
A total of 86 vertebrae of 6 patients with early-onset scoliosis and 3 patients with congenital scoliosis, age ranged between 7.8 and 12.5 years, who had both thoracic spine CT scan and low-dose stereoradiography of the spine were included. 3D reconstructions of CT and low-dose stereoradiography were generated. Using previously validated image processing techniques, vertebral anterior (A), posterior (P), left (L), and right (R) heights, superior and inferior endplates depth and width (S-D, I-D, S-W, I-W) were measured on the CTs' 3D reconstructions and were compared with the same parameters measured on low-dose stereoradiography reconstructions using a postprocessing custom code. The agreement between the 2 techniques in measurement of the vertebral morphology was assessed using the Bland-Altman plots.
Results:
No significant difference was observed in the A, P, L, R, S-W, and I-W between the stereoradiography and CT measurements (P>0.05). S-D and I-D were significantly greater in low-dose stereoradiography measurements P<0.05. Bland-Altman plots showed an agreement between the stereoradiography and CT techniques in vertebral height measurements (A, P, L, R); however, larger measurement bias and greater limits of agreement in S-D, I-D, S-W, and I-W measurements were shown.
Conclusions:
In vivo measurements of the vertebral heights using low-dose stereoradiography 3D reconstructions were comparable with CT measurements except for the superior and inferior vertebral depths. The low-dose stereoradiography imaging modality and the postprocessing platform can be used for assessment of the vertebral heights and monitoring asymmetric growth in patients undergoing growth-sparing treatment except for the vertebral levels with congenital deformities.
Level Of Evidence:
Level IV.
Related Concept Videos
04:48Application of Deep Learning-Based Medical Image Segmentation via Orbital Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
10:35Precision Measurements and Parametric Models of Vertebral Endplates
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
06:56Tree Core Analysis with X-ray Computed Tomography