Sources of Error in Office Blood Pressure Measurement
Roy N Morcos1, Kimbroe J Carter2, Frank Castro2
1From the Department of Family and Community Medicine, Northeast Ohio Medical University, Rootstown, OH (RNM, SK); Department of Family Medicine Residency Program, St. Elizabeth Boardman, Mercy Health (RNM); Medical Decision Making Society of Youngstown Ohio, c/o Jeghers Medical Index, St. Elizabeth Youngstown Hospital, Youngstown, OH (FC, KJC); Family Medicine Residency Program, St. Elizabeth Boardman, Mercy Health (SK, DS); Department of Pathology, Northeast Ohio Medical University, Rootstown, OH (KJC). Roy_Morcos@mercy.com.
Purposes:
To evaluate 2 commonly overlooked sources of error in measuring blood pressure (BP) in the office, improper patient positioning and frequency of terminal digit bias (TDB) using manual and automated (BP) devices.
Methods:
BPs recorded by 3 nurses using manual and automated devices were analyzed for TDB. In the next part of the study, 294 patients were recruited and tested with each patient's BP measured twice in the table position and compared with BP measured in the chair position. To eliminate concern for position sequence, a randomized controlled trial was initially conducted.
Results:
Significant TDB for the digit zero was identified in BPs measured by all nurses using a manual device. No such bias was identified for any nurse when measuring BP with an automated device. For the positional study, the randomized controlled study showed no significant sequencing effect therefore the sequence of table then chair BP measurements was adopted. Significant BP lowering was observed in 128 patients (42.7%) in the chair compared with the table position. Misclassification of prehypertension and hypertension would have occurred in 15.3% and 16% of patients, respectively, when BP was recorded in the table instead of the chair position.
Conclusions:
Significant TDB was identified for all nurses when using a manual but not an automated device. Patient positioning on the examination table resulted in elevations of systolic and diastolic BPs.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Noninvasive Blood Pressure Measurement Techniques
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Blood Pressure Measurement
The term blood pressure (BP) describes the lateral pressures produced by blood upon the vessel walls. BP is a vital sign obtained routinely in hospital and outpatient settings, and is one of the most common medical assessments performed around the world. It can be determined directly with the intra-arterial catheter or by an indirect method, which is a non-invasive, safe, easily reproducible, and thus most...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.