Optimization of radiation dose for CT detection of lytic and sclerotic bone lesions: a phantom study
J Greffier1, J Frandon2, F Pereira2
1Service d'imagerie medicale, CHU Nimes, Univ Montpellier, Medical Imaging Group Nimes, EA 2415, Bd Prof Robert Debré, 30029, Nîmes Cedex 9, France. joel.greffier@chu-nimes.fr.
Objectives:
To determine the best compromise between low radiation dose and suitable image quality for the detection of lytic and sclerotic bone lesions of the lumbar spine and pelvis.
Methods:
A phantom was scanned using the routine protocol (STD, 13 mGy) and six decreasing dose levels. Raw data were reconstructed using level 3 of iterative reconstruction (IR3) with 1-mm slice thickness for the STD protocol and highest IR levels with 3-mm slice thickness for the others. CTDIvol was used for radiation dose assessment. Quantitative criteria (noise power spectrum [NPS], task-based transfer function [TTF], and the detectability index [d']), as well as qualitative analysis, were used to compare protocols. NPS and TTF were computed using specific software (imQuest). d' was computed for two imaging tasks: lytic and sclerotic bone lesions. A subjective analysis was performed to validate the image quality obtained on the anthropomorphic phantom with the different dose values.
Results:
Similar d' values were found for CTDIvol from 3 to 4 mGy with IR4 and from 1 to 2 mGy for IR5 compared with d' values using the STD protocol. Image quality was validated subjectively for IR4 but rejected for IR5 (image smoothing). Finally, for the same d', the dose was reduced by 74% compared with the STD protocol, with the CTDIvol being 3.4 mGy for the lumbar spine and for the pelvis.
Conclusion:
A dose level as low as 3.4 mGy, in association with high levels of IR, provides suitable image quality for the detection of lytic and sclerotic bone lesions of the lumbar spine and pelvis.
Key Points:
• A CTDI volof 3.4 mGy, in association with high iterative reconstruction level, provides suitable image quality for the detection of lytic and sclerotic bone lesions, both at objective and subjective analysis. • Compared with the standard protocol, radiation dose can be reduced up to 74% for the lumbar spine and pelvis. • A task-based image quality assessment using the detectability index represents an objective method for the assessment of image quality and bridges the gap between complex physical metrics and subjective image analysis.
More Related Videos
Related Concept Videos
08:04Monitoring Tumor Metastases and Osteolytic Lesions with Bioluminescence and Micro CT Imaging
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
09:26Preparation and Using Phantom Lesions to Practice Fine Needle Aspiration Biopsies
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
10:25Brain Infarct Segmentation and Registration on MRI or CT for Lesion-symptom Mapping
08:17Low-Dose Gamma Radiation Sterilization for Decellularized Tracheal Grafts