Related Experiment Videos
Evaluation of several film/screen combinations for megavoltage radiation therapy treatment field verification
I J Das1, E C McCullough, F Weber
1Division of Radiation Oncology, Mayo Clinic/Foundation, Rochester, MN 55905.
Abstract:
The introduction of a lower cost alternative to a film commonly used for megavoltage radiation therapy port films led to a comparison of two films (Kodak X-TL and X-OMAT G) in four metal screen cassette configurations. In addition to determination of H-D curves and point spread functions, images of humanoid phantom sections were obtained and evaluated for clinical "usefulness." The X-OMAT G film-screen combinations compared favorably with the X-TL film combinations with respect to the H-D Curves and point spread functions but there was a slight preference for X-TL film. There were no differences noted whether the film was used in the standard "ready pack" or naked in the cassettes.