Related Experiment Video
Updated: Jan 19, 2026

Sound Source Localization Testing in Single-sided Deafness Following Bone Conduction Intervention
Published on: December 20, 2024
Data Logging Evidence of Cochlear Implant Use in Single-Sided and Bilateral Deafness
Ann-Kathrin Rauch1, Stephanie Kagermann2, Thomas Wesarg2
1Department of Otorhinolaryngology, Head and Neck Surgery, ENT Clinic, Medical Centre, University of Freiburg, Freiburg, Germany, ann-kathrin.rauch@uniklinik-freiburg.de.
Purpose:
This study analyses data logs in order to investigate the usage pattern of cochlear implant (CI) recipients with single-sided deafness (SSD-CI) and bilaterally deaf, uni- or bilaterally implanted CI recipients (Uni-CI and Bil-CI). Data logging is available from SCAN, an automated auditory scene classifier which categorizes auditory input into 6 listening environments.
Methods:
CI usage data were retrospectively available from data logs of 206 CI recipients using the Nucleus 6 system obtained between January 2013 and June 2015. For all recipients, we analysed time on air and time spent in the listening environments. For statistical analysis, we matched the CI recipients according to age and duration of CI experience and classified them into 4 age groups.
Results:
SSD-CI showed a similar time on air compared to Uni- and Bil-CI. Usage behaviour of SSD-CI was comparable to Uni- and Bil-CI regarding exposure to music, speech in quiet and speech in noise. With increasing age, exposure to quiet increased and exposure to music decreased across all CI recipient groups in relation to time on air.
Conclusion:
In total, the CI usage pattern of SSD-CI is comparable for the majority of listening environments and age groups to that of Uni- and Bil-CI. The results of our study show that SSD-CI benefit equally from CI implantation.
Related Concept Videos
04:32Sound Source Localization Testing in Single-sided Deafness Following Bone Conduction Intervention
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
08:06Robotic Cochlear Implantation for Direct Cochlear Access
09:56Cochlear Implantation in the Guinea Pig
06:16The Miniature Pig: A Large Animal Model for Cochlear Implant Research
06:54Performing Repeated Intraoperative Impedance Telemetry Measurements during Cochlear Implantation