Background:
Adverse reactions to metal debris after total hip arthroplasty are an increasing concern. Metal debris is released from the metal-on-metal articulation itself as well as from the head-neck taper. There is also a risk of corrosion at the neck-stem junction. Some stems were recalled because of high revision rates as a result of metal debris from the modular junction. We conducted a retrospective analysis of the 3-year to 8-year clinical and radiographic outcomes of a consecutive cohort of patients who underwent cementless total hip arthroplasty using a modular hip prosthesis with ceramic-on-ceramic bearings.
Methods:
We retrospectively evaluated 104 patients who had undergone 119 primary ceramic-on-ceramic total hip arthroplasties using modular neck femoral components between December 2006 and November 2012. All hips were implanted with a titanium alloy stem and titanium alloy neck. Clinical outcomes were evaluated using the Japanese Orthopaedic Association hip score. Prosthesis survival was calculated according to the Kaplan-Meier method considering any reason for revision as failure. Postoperative magnetic resonance imaging was performed and graded using a previously described pseudotumor grading system. Mean follow-up was 5.9 years (range: 3 to 8 years).
Results:
The mean age of the patients at the time of the index arthroplasty was 59.0 years. Mean hip score improved from 45.9 (range: 9 to 90) preoperatively to 91.5 (range: 65 to 100) at the final follow-up. The overall implant survival rate was 97.3% at 8 years when revision for any reason was used as the endpoint. Abnormalities on magnetic resonance imaging were seen in 16 of 91 hips (17.6%). All abnormalities were cystic thin-walled lesions located in the peritrochanteric area. There was no revision operation due to adverse reactions to metal debris.
Conclusions:
A modular neck femoral prosthesis with ceramic-on-ceramic bearings offers satisfactory 3-year to 8-year results. However, abnormalities on magnetic resonance imaging were seen in one-fifth of the patients, although the clinical significance is unknown.
Related Concept Videos
08:29Multi-material Ceramic-Based Components – Additive Manufacturing of Black-and-white Zirconia Components by Thermoplastic 3D-Printing (CerAM - T3DP)
08:43Fused Filament Fabrication (FFF) of Metal-Ceramic Components
06:53Additive Manufacturing of Functionally Graded Ceramic Materials by Stereolithography
09:31Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
05:42Orthopedic Robot-Assisted Femoral Neck System in the Treatment of Femoral Neck Fracture
Ceramic-matrix Composite Materials and Their Bending Properties
Bones are composites, made of a ceramic matrix and polymer fiber reinforcements. The ceramic contributes compressive strength, and the polymer provides tensile and flexural strength. By combining ceramic and polymer materials in different amounts, the body can create unique materials tailored for a specific application. As biomedical engineers, having the ability...