Related Experiment Video
Updated: Jan 19, 2026
Cardiac MRI: Principle and Imaging
Published on: April 30, 2023
Unexpected magnetic resonance imaging burn injuries from jogging pants
Hiroyuki Tokue1, Azusa Tokue1, Yoshito Tsushima1
1Department of Diagnostic and Interventional Radiology, Gunma University Hospital, 3-39-22 Showa-machi, Maebashi, Gunma 371-8511, Japan.
Abstract:
During magnetic resonance imaging, thermal burns are a potential cause of injury. We experienced a rare case of magnetic resonance imaging burn injury caused by jogging pants. The list of materials of the jogging pants did not indicate any metal. However, light, thin metal fibers were used. Changing to suitable magnetic resonance imaging-compatible clothing is important before undergoing magnetic resonance imaging. To minimize the risk of magnetic resonance imaging burns, it is necessary to increase staff awareness and communicate and comply with pretreatment safety checks.
Related Concept Videos
Cardiac Magnetic Resonance Imaging
In this video, high field, small-bore magnetic resonance imaging (MRI) with physiological monitoring is demonstrated to acquire gated cine loops of the murine cardiovascular system. This procedure provides a basis for assessing left-ventricular function, visualizing vascular networks, and quantifying motion of organs due to respiration. Comparable small animal...
Burn Injuries
The damage results in the death of skin cells, which can lead to a massive loss of fluid. Dehydration, electrolyte imbalance, and renal and circulatory failure follow, which can be fatal. Burn patients are treated with intravenous fluids to offset...
09:08Multiple-mouse Neuroanatomical Magnetic Resonance Imaging
fMRI: Functional Magnetic Resonance Imaging
Magnetic Resonance Imaging
09:30Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease