TonoVet versus Tonopen in a high intraocular pressure monkey model
Shu Tu1, Kang Li1, Xiaohu Ding1
1State Key Laboratory of Ophthalmology, Guangdong Provincial Key Laboratory of Ophthalmology and Visual Science, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, China.
Purpose:
The purpose of this study is to evaluate the consistency of and deviation in intraocular pressure (IOP) measurements obtained by the TonoVet rebound tonometer and the Tonopen applanation tonometer in a primate model.
Methods:
Twenty-four-hour IOPs (nine time points) were recorded in ten monkeys with normal IOP and eight monkeys with chronic high IOP (one eye was randomly selected for measurement in each animal) using a Tonopen and TonoVet device. Measurements obtained using both handheld devices were first compared in the healthy control group (90 readings). The monkeys with chronic ocular hypertension (COHT, 72 IOP readings) were divided into three subgroups according to the level of IOP. The consistency of and deviations in the measurements were analyzed using Bland-Altman plots, linear regression, and two-tailed Student t tests.
Results:
In monkeys with normal IOP, the two devices produced similar IOP readings (mean IOP deviation, 0.06 ± 2.08 mmHg, p = 0.761), with 56.67% of the deviation between -1 mmHg and 1 mmHg and 91.12% between -3 mmHg and 3 mmHg. However, in the animal model group (23-60 mmHg), the readings obtained by the TonoVet tonometer were higher than those obtained by the Tonopen tonometer (mean deviation, 13.76 ± 9.19 mmHg); furthermore, 75.68% of the TonoVet measurements deviated by ± 5 mmHg from the Tonopen measurements.
Conclusions:
In animals with normal IOP, the TonoVet and Tonopen tonometers produced consistent measurements. However, in a monkey model of chronic high IOP, the measurements obtained by these tonometers were inconsistent, with higher IOPs associated with larger measurement errors. Therefore, it is necessary to be aware that among different tonometers, there may be systemic errors and deviations in IOP measurements. These findings should facilitate efforts to obtain more accurate individualized diagnoses and prevent the utilization of misleading IOP values.
More Related Videos
09:46A Pressure Injection System for Investigating the Neuropharmacology of Information Processing in Awake Behaving Macaque Monkey Cortex
Published on: March 14, 2016
08:55Translaminar Autonomous System Model for the Modulation of Intraocular and Intracranial Pressure in Human Donor Posterior Segments
Published on: April 24, 2020
Related Concept Videos
Induction of Elevated Intraocular Pressure in a Rat Model
09:46A Pressure Injection System for Investigating the Neuropharmacology of Information Processing in Awake Behaving Macaque Monkey Cortex
08:55Translaminar Autonomous System Model for the Modulation of Intraocular and Intracranial Pressure in Human Donor Posterior Segments
07:37A Mouse Model of Retinal Ischemia-Reperfusion Injury Through Elevation of Intraocular Pressure
07:26Synthesis and Microdiffraction at Extreme Pressures and Temperatures
10:02Use of Rabbit Eyes in Pharmacokinetic Studies of Intraocular Drugs