Related Experiment Video
Updated: Jan 19, 2026

Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
A Novel Algorithm in Radiation Dosimetry of Regular and Irregular Treatment Fields
Fatemeh Seif1, Nayyer Mostafavi2, Mohammadreza Bayatiyani3
1Department of Radiotherapy and Oncology, Ayatollah Khansari Hospital, Arak, Iran.
Background:
The aim of this study was to design an algorithm for the calculation of monitor unit (MU) in a short time and high precision for different radiotherapy (RT) fields.
Materials And Methods:
The algorithm for calculating MU for the stated patients was designed in MATLAB software. To investigate the efficiency of this algorithm, 11 regular chest fields with the sizes of 7 cm × 7 cm up to 17 cm × 17 cm were considered, and the obtained MUs were compared with MUs of 13 patients which were calculated with a "hand calculation" which is used in some RT centers for the aforementioned fields.
Results:
The maximum percentage of calculation errors of regular fields at the depths of 4 and 10 cm were 1 and 0.8, respectively. The maximum and minimum percentage of calculation errors in irregular fields was 3 and 0.9, respectively. Furthermore, the maximum and minimum errors were 8.8 and 0.14, respectively. In addition, relative percentages of the MUs for irregular fields of chest and supraclavicular were 1.63 and 1.01, respectively.
Conclusion:
Calculation of MUs is suggested to be performed with the novel proposed algorithm, due to reduce the treatment time, and also provide high accuracy and precision compared to hand calculation.
More Related Videos
Related Concept Videos
08:25Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
06:51Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Areas Within Irregular Boundaries
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
05:56Construction of a Compact Low-Cost Radiation Shield for Air-Temperature Sensors in Ecological Field Studies