Influence of Cone Beam Computed Tomography Settings on Implant Artifact Production: Zirconia and Titanium
Purpose:
This in vitro study assessed the artifact production related to titanium and zirconia implants in cone beam computed tomography (CBCT) and compared the effect of different protocol settings on image quality for both materials.
Materials And Methods:
A titanium implant and a zirconia implant were placed in a dry mandible. CBCT scans were obtained separately for each implant using the ProMax 3D (Planmeca) unit; 20 protocols were tested with varying kilovoltage (70 to 90 kVp) and resolution (high and low), and with and without a metal artifact reduction tool. Standard deviation and contrast-noise ratio were calculated in regions of interest adjacent and distant to the implant.
Results:
The zirconia produced more artifacts and its images were more affected by the different protocols than titanium. High kVps and an activated metal artifact reduction tool decreased the standard deviation values related to both implants. Activation of the metal artifact reduction tool also increased contrast-noise ratio values for both implants, whereas increasing kVp improved them only on titanium images. The standard deviation and contrast-noise ratio were not affected by resolution.
Conclusion:
The zirconia implant generated more image artifacts than the titanium implant. Increasing kVp and the metal artifact reduction tool are efficient in decreasing the CBCT artifacts for both implants, whereas resolution does not affect their production.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
10:23Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
05:49Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor