Related Experiment Video
Updated: Jan 19, 2026

Translational Rabbit Model of Chronic Cardiac Pacing
Published on: January 6, 2023
Effectiveness of first versus successive antitachycardia pacing attempts: predictors and clinical consequences
Javier Jiménez-Candil1, Olga Durán2, Jean Núñez2
1Arrhythmias Unit, Cardiology Department, IBSAL-University Hospital, CIVER-CV, Universidad de Salamanca, Paseo de San Vicente, 58-182, 37007, Salamanca, Spain. jimenezcandil@secardiologia.es.
Purpose:
Antitachycardia pacing (ATP) terminates the majority (but not all) of slow ventricular tachycardias (S-VT) with a cycle length (CL) > 320 ms. Usually, several ATP therapies are programmed in the S-VT zone. Our objective is to analyse the ATP effectiveness, comparing the first ATP attempt (ATP-1) to the second (ATP-2) and third (ATP-3) attempts.
Methods:
We studied 556 S-VT (CL = 354 ± 18). ATP programming was standardized and included three bursts of 15 pulses at 91% of VT CL.
Results:
ATP effectiveness declined from ATP-1 (436/556: 78%) compared to ATP-2 (24/103: 23%) and ATP-3 (10/79: 13%) (p < 0.01) for all comparisons. The percentage of variation of RR intervals (P-RR, %) was higher prior to effective ATP-1 (2.73 ± 1.45 vs. 1.23 ± 0.9; p < 0.001). After an ineffective ATP-1, the P-RR decreased dramatically, with no differences between episodes terminated or not at ATP-2 (0.6 ± 0.14 vs. 0.44 ± 0.16; p = 0.6) or ATP-3 (0.54 ± 0.15 vs. 0.52 ± 0.14; p = 0.7). The post-pacing interval-CL difference (PPI-TCLd) after an unsuccessful ATP-1 was shorter in episodes terminating at ATP-2 or ATP-3 (180 ± 24 vs. 211 ± 15 ms; p < 0.001). Several independent predictors of ATP efficacy were found, as follows: (a) ATP-1: P-RR, % (OR = 7.3; p < 0.001), beta-blockers (OR = 4.1; p < 0.001) and QRS ≥ 120 ms (OR = 0.3; p < 0.001); (b) ATP-2: PPI-TCLd, ms (OR = 0.94; p = 0.001) and QRS ≥ 120 ms (OR = 0.6; p = 0.04); (c) ATP-3: PPI-TCLd, ms (OR = 0.93; p = 0.009).
Conclusions:
The effectiveness of ATP is mainly due to ATP-1. The regularization of RR intervals after ineffective ATP-1 underlies the lower efficacy of successive attempts. Shorter PPI-TCLd is associated with higher effectiveness of ATP-2 and ATP-3. Since a duration of QRS ≥ 120 ms predicts a longer PPI-TCLd, patients with wide QRS complexes have less effective ATP-2 and APT-3.
Related Concept Videos
06:14Translational Rabbit Model of Chronic Cardiac Pacing
Ecological Succession
Timing and Consequences on Behavior
Humans, however, can respond to delayed reinforcers. We often make decisions between immediate small rewards and delayed larger rewards. This ability to delay gratification is a significant...
11:06Structural Characterization of Mannan Cell Wall Polysaccharides in Plants Using PACE
08:05Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
05:12Transesophageal Atrial Burst Pacing for Atrial Fibrillation Induction in Rats