Measurement of isocenter alignment accuracy and image distortion of an 0.35 T MR-Linac system
1Department of Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), INF 280, 69120 Heidelberg, Germany. Faculty of Physics and Astronomy, University of Heidelberg, Heidelberg, Germany. National Center for Radiation Research in Oncology (NCRO), Heidelberg Institute for Radiation Oncology (HIRO), Heidelberg, Germany. Author to whom correspondence should be addressed.
Abstract:
For hybrid devices combining magnetic resonance (MR) imaging and a linac for radiation treatment, the isocenter accuracy as well as image distortions have to be checked. This study presents a new phantom to investigate MR-Linacs in a single measurement in terms of (i) isocentricity of the irradiation and (ii) alignment of the irradiation and imaging isocenter relative to each other using polymer dosimetry gel as well as (iii) 3-dimensional (3D) geometric MR image distortions. The evaluation of the irradiated gel was performed immediately after irradiation with the imaging component of the 0.35 T MR-Linac using a T2-weighted turbo spin-echo sequence. Eight plastic grid sheets within the phantom allow for measurement of geometric distortions in 3D by comparing the positions of the grid intersections (control points) within the MR-image with their nominal position obtained from a CT-scan. The distance of irradiation and imaging isocenter in 3D was found to be (0.8 ± 0.9) mm for measurements with 32 image acquisitions. The mean distortion over the whole phantom was (0.60 ± 0.28) mm and 99.8% of the evaluated control points had distortions below 1.5 mm. These geometrical uncertainties have to be considered by additional safety margins.
Related Concept Videos
06:51Measuring the Complete-arch Distortion of an Optical Dental Impression
Uncertainty in Measurement: Accuracy and Precision
07:57Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
11:56Fluorescence Imaging with One-nanometer Accuracy (FIONA)
07:03Medical-grade Sterilizable Target for Fluid-immersed Fetoscope Optical Distortion Calibration
Improving Translational Accuracy