Multilevel qualification of a large set of blood gas analyzers: Which performance goals?
Nils Kuster1, Anne Sophie Bargnoux1, Stéphanie Badiou1
1Department of Biochemistry and Hormonology, CHU de Montpellier, France; PhyMedExp, University of Montpellier, INSERM, CNRS, France.
Background:
Blood gas analyzers are frequently installed as point of care devices and thus allow rapid decision making. Few data are available regarding analytical performance of large sets of BGA. We aimed at evaluating 22 ABL 90 Flex Plus analyzers intended to be deployed. The evaluation was performed at the device level and at the entire set level to characterize the quality of measurements but also to ensure consistency across the devices deployed in the hospital.
Methods:
Imprecision and total error were assessed for pH, pCO2, pO2, sodium, potassium, ionized calcium, glucose, lactate and oximetry parameters. Imprecision at the hospital level including between device variability was also evaluated. One of the two analyzers used in the central laboratory was correlated with a GEM Premier 4000 and a Cobas b221 analyzers. Thereafter, we tested sequentially the 20 instruments intended to be deployed in care service in comparison with the reference device.
Results:
Heterogeneity of analytical performance across the different analyzers was low, allowing to consider the whole set as a unique analyzer. The total error was in line with performance goals. Analytical performance of the analyzers was found suitable for use in clinical practice.
Conclusions:
Our study is an example of the qualification of a set of point and underscores 1)The need for a unified qualification scheme when multiple analyzers are deployed simultaneously 2) analytical performance goals compatible with clinical use and the state of the art for all parameters.
Related Concept Videos
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Data Validation
Key parameters for method validation include:
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Mass Analyzers: Overview
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...