Inappropriate noise detection in Tendril family pacing leads
Dennis W X Zhu1,2,3, Matthew M Chu1, Andrew J Zager1,2
1Department of Cardiology, Regions Hospital, St Paul, Minnesota.
Noise oversensing occurred in 6.2% of Abbott Tendril leads, but not in Medtronic leads connected to similar generators. Most affected leads were managed conservatively, with lead malfunction identified in a small percentage.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Noise oversensing is a potential issue in cardiac implantable electronic devices.
- Abbott Tendril leads have been associated with oversensing events.
Purpose of the Study:
- To investigate the incidence and characteristics of noise oversensing in Abbott Tendril leads.
- To compare the performance of Tendril leads with Medtronic CapSureFix leads in Abbott generators.
Main Methods:
- Retrospective analysis of patients with Abbott Tendril leads (models 1688, 1788, 1888, 2088).
- Prospective enrollment of patients with Medtronic CapSureFix leads (models 4076, 5076) connected to Abbott generators.
- Evaluation of lead impedance, symptoms, interventions, and explanted lead analysis.
Main Results:
- Noise oversensing was detected in 6.2% of 1063 Tendril leads over 3.9 years.
- Most affected Tendril leads had normal impedance; 1.5% required replacement.
- Lead-to-device abrasion and insulation breach were found in explanted Tendril leads.
- No noise oversensing was observed in 145 Medtronic CapSureFix leads connected to Abbott generators over 3.6 years.
Conclusions:
- Noise oversensing is relatively common with Abbott Tendril leads.
- Medtronic leads did not exhibit noise oversensing when connected to Abbott generators.
- Conservative management is often feasible for asymptomatic patients with Tendril lead noise oversensing.
Related Concept Videos
Cardiovascular System Abnormal Findings II: Auscultation
Abnormal Heart Sounds
Gallops:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias II: Classification of Tachyarrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities: