Agreement between frontal and occipital regional cerebral oxygen saturation in infants during surgery and general anesthesia an observational study
Related Experiment Video
Updated: Jan 19, 2026

How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Pether Jildenstål1,2,3, Johan Sandin4, Margareta WarrènStomberg1
1Institute of Health and Care Sciences, Sahlgrenska academy, University of Gothenburg, Gothenburg, Sweden.
Background:
Advances in perioperative pediatric care have resulted in an increased number of procedures requiring anesthesia. During anesthesia and surgery, the patient is subjected to factors that affect the circulatory homeostasis, which can influence oxygenation of the brain. Near-infrared spectroscopy (NIRS) is an easy applicable noninvasive method for monitoring of regional tissue oxygenation (rScO₂%). Alternate placements for NIRS have been investigated; however, no alternative cranial placements have been explored.
Aim:
To evaluate the agreement between frontal and occipital recordings of rScO₂% in infants using INVOSTM during surgery and general anesthesia.
Method:
A standard frontal monitoring of rScO₂% with NIRS was compared with occipital rScO₂% measurements in fifteen children at an age <1 year, ASA 1-2, undergoing cleft lip and/or palate surgery during general anesthesia with sevoflurane. An agreement analysis was performed according to Bland and Altman.
Results:
Mean values of frontal and occipital rScO₂% at baseline were largely similar (70.7 ± 4.77% and 69.40 ± 5.04%, respectively). In the majority of the patients, the frontal and occipital recordings of rScO2 changed in parallel. There was a moderate positive correlation between frontal and occipital rScO₂% INVOS™ readings (rho[ρ]: 0.513, P < .01). The difference between frontal and occipital rScO₂ ranged from -31 to 28 with a mean difference (bias) of -0.15%. The 95% limit of agreement was -18.04%-17.74%. The error between frontal and occipital rScO₂ recordings was 23%.
Conclusion:
The agreement between frontal and occipital recordings of brain rScO₂% in infants using INVOSTM during surgery and general anesthesia was acceptable. In surgical procedures where the frontal region of the head is not available for monitoring, occipital recordings of rScO₂% could be an option for monitoring.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Stages of General Anesthesia