Related Experiment Video
Updated: Jan 19, 2026

Author Spotlight: Improved Localization and Monitoring of Coronary Flow Reserve Using Modified PLAX View in Mice
Published on: August 25, 2023
Diagnostic performance of a fast non-invasive fractional flow reserve derived from coronary CT angiography: an
1Department of Radiology, Beijing Anzhen Hospital, Capital Medical University, 2 Anzhen Rd Chaoyang District, Beijing, 100029, China.
Aim:
To validate the computed tomography (CT)-derived fractional flow reserve (FFRCT) that was computed by new, fast, automatic software and to compare the diagnostic accuracy between FFRCT and stenosis diagnosed at coronary CT angiography (CCTA).
Materials And Methods:
A total of 110 patients (76 males, 59±9 years) and 125 vessels underwent CCTA. FFRCT was computed by fast automatic software and compared with invasive FFR. The diagnostic performance between FFRCT and CCTA-diagnosed stenosis were compared on the per-patient and per-vessel level.
Results:
The computational time of FFRCT is 10±5 minutes (averaged over 125 vessels). The FFRCT has a good correlation with invasive FFR (r=0.59, p<0.0001) with a small bias of -0.02 (-0.26-0.23). The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of FFRCT were 76.5, 89.5, 89.7, and 76.1% on a vessel level. The area under the receiver operating characteristic curve of FFRCT was higher than CCTA-diagnosed stenosis (0.82 versus 0.72, P=0.034).
Conclusion:
The computation of FFRCT is possible and reliable when using the new, fast, automatic software first employed in the present clinical study. The FFRCT has a good correlation with invasive FFR and provides better diagnostic performance than CCTA-diagnosed stenosis.
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System V: CT