Related Experiment Video
Updated: Jan 19, 2026

Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025
Characterization and Mechanical Proprieties of New TiMo Alloys Used for Medical Applications
Andrei Victor Sandu1,2, Madalina Simona Baltatu3, Marcin Nabialek4
1"Gheorghe Asachi" Technical University of Iasi, Faculty of Materials Science and Engineering, 700050 Iasi, Romania. sav@tuiasi.ro.
Abstract:
Ti-based alloys are accessible for use in the human body due to their good mechanical properties, corrosion resistance, and biocompatibilities. These main properties of alloys are important criteria for choosing biomedical implants for human bones or for other kinds of applications in general medicine. This paper presents a comparison of four new Ti-based alloys desired to satisfy various requirements for biomedical implants. The materials were prepared with recipes for two new system alloys, TiMoZrTa (TMZT) and TiMoSi (TMS), alloys with nontoxic elements. The presented research contains microstructure images, indentation tests, Vickers hardness, XRD, and corrosion resistance, showing better characteristics than most commercial products used as implants (Young's modulus closer to the human bone).