One Year Clinical Experience of the First Commercial Hybrid Closed-Loop System
Rayhan A Lal1,2,3, Marina Basina4,3, David M Maahs2,3
1Division of Endocrinology, Department of Medicine, Stanford University School of Medicine, Stanford, CA inforay@stanford.edu.
Objective:
In September 2016, the U.S. Food and Drug Administration approved the Medtronic 670G "hybrid" closed-loop system. In Auto Mode, this system automatically controls basal insulin delivery based on continuous glucose monitoring data but requires users to enter carbohydrates and blood glucose for boluses. To track real-world experience with this first commercial closed-loop device, we prospectively followed pediatric and adult patients starting the 670G system.
Research Design And Methods:
This was a 1-year prospective observational study of patients with type 1 diabetes starting the 670G system between May 2017 and May 2018 in clinic.
Results:
Of the total of 84 patients who received 670G and consented, 5 never returned for follow-up, with 79 (aged 9-61 years) providing data at 1 week and 3, 6, 9, and/or 12 months after Auto Mode initiation. For the 86% (68 out of 79) with 1-week data, 99% (67 out of 68) successfully started. By 3 months, at least 28% (22 out of 79) had stopped using Auto Mode; at 6 months, 34% (27 out of 79); at 9 months, 35% (28 out of 79); and by 12 months, 33% (26 out of 79). The primary reason for continuing Auto Mode was desire for increased time in range. Reasons for discontinuation included sensor issues in 62% (16 out of 26), problems obtaining supplies in 12% (3 out of 26), hypoglycemia fear in 12% (3 out of 26), multiple daily injection preference in 8% (2 out of 26), and sports in 8% (2 out of 26). At all visits, there was a significant correlation between hemoglobin A1c (HbA1c) and Auto Mode utilization.
Conclusions:
While Auto Mode utilization correlates with improved glycemic control, a focus on usability and human factors is necessary to ensure use of Auto Mode. Alarms and sensor calibration are a major patient concern, which future technology should alleviate.
More Related Videos
06:29Wireless Telemetry Device Implantation in a Fontan Ovine Model for Continuous and Long-Term Hemodynamic Monitoring
Published on: May 2, 2025
11:02The Use of an Automated System GreenFeed to Monitor Enteric Methane and Carbon Dioxide Emissions from Ruminant Animals
Published on: September 7, 2015
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...