Hazard Group 3 agent decontamination using hydrogen peroxide vapour in a class III microbiological safety cabinet
T Pottage1, S Lewis2, A Lansley2
1Biosafety, Air and Water Microbiology Group, National Infection Service, Public Health England, Salisbury, UK.
Aims:
This study investigated the efficacy of hydrogen peroxide vapour (HPV) at inactivating hazard group 3 bacteria that have been presented dried from their growth medium to present a realistic challenge.
Methods And Results:
Hydrogen peroxide vapour technology (Bioquell) was used to decontaminate a class III microbiological safety cabinet containing biological indicators (BIs) made by drying standard working suspensions of the following agents: Bacillus anthracis (Ames) spores, Brucella abortus (strain S99), Burkholderia pseudomallei (NCTC 12939), Escherichia coli O157 ST11 (NCTC 12079), Mycobacterium tuberculosis (strain H37Rv) and Yersinia pestis (strain CO92) on stainless steel coupons. Extended cycles were used to expose the agents for 90 min. The HPV cycle completely inactivated B. anthracis spores, B. abortus, B. pseudomallei, E. coli O157 and Y. pestis when BIs were processed using quantitative and qualitative methods. Whilst M. tuberculosis was not completely inactivated, it was reduced by 4 log10 from a starting concentration of 106 colony-forming units.
Conclusions:
This study demonstrates that HPV is able to inactivate a range of HG3 agents at high concentrations with associated organic matter, but M. tuberculosis showed increased resistance to the process.
Significance And Impact Of The Study:
This publication demonstrates that HPV can inactivate HG3 agents that have an organic load associated with them. It also shows that M. tuberculosis has higher resistance to HPV than other agents. This shows that an appropriate BI to represent the agent of interest should be chosen to demonstrate a decontamination is successful.
Related Concept Videos
Chemical Agents for Microbial Control
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...