Iodine quantification and detectability thresholds among major dual-energy CT platforms
Ross Edward Taylor1, Pamela Mager1, Nam C Yu1
1Department of Radiology, Houston Methodist Hospital, Houston, TX.
Objectives:
To estimate the minimum detectable iodine concentration on multiple dual-energy CT (DECT) platforms.
Methods And Materials:
A phantom containing iodine concentrations ranging from 0 to 10 mg ml-1 was scanned with five dual-energy platforms (two rapid kilo volt switching (r-kVs), one dual source (DS), one sequential acquisition and one split-filter). Serial dilutions of 300 mg ml-1 iodinated contrast material were used to generate concentrations below 2 mg ml-1. Iodine density and virtual monoenergetic images were reviewed by three radiologists to determine the minimum visually detectable iodine concentration. Contrast-to-noise ratios (CNRs) were calculated.
Results:
1 mg mL-1 (~0.8 mg mL-1 corrected) was the minimum visually detectable concentration among the platforms and could be seen by all readers on the third-generation r-kVs and DS platforms.
Conclusions:
At low concentrations, CNR for monoenergetic images was highest on the DS platform and lowest in the sequential acquisition and split-filter platforms.
Advances In Knowledge:
The results of this study corroborate previous in vivo estimates of iodine detection limits at DECT and provide a comparison for the performance of different DECT platforms at low iodine concentrations in vitro.
More Related Videos
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...