Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Google fit smartphone application or Gt3X Actigraph: Which is better for detecting the stepping activity of
Janaine Cunha Polese1, Giselle Silva E Faria2, Giane Amorim Ribeiro-Samora3
1Department of Physical Therapy, Faculdade Ciências Médicas de Minas Gerais, Belo Horizonte, Minas Gerais, Brazil; NeuroGroup, Department of Physical Therapy, Universidade Federal de Minas Gerais, Belo Horizonte, Minas Gerais, Brazil.
Objective:
To examine the validity of the GT3X® ActiGraph accelerometer and the Google Fit® smartphone application in estimating stepping activity in people with chronic stroke.
Methods:
Thirty-seven stroke survivors walked along a straight, 10 metre hallway over 5 min at their fastest speeds, wearing the GT3X® ActiGraph accelerometer and a smartphone on their paretic lower limb. The criterion-standard measurement made was the actual number of steps, counted by a trained examiner.
Results:
The mean estimated steps measured by the GT3X® ActiGraph and Google Fit® respectively were 276.7 ± 97.6 and 481.0 ± 119.8; that calculated from the examiner's measurements was 472.0 ± 93.9. Statistically significant associations were found between the actual steps and those estimated by the GT3X® ActiGraph (r = 0.56; p < 0.001) and Google Fit® (r = 0.89; p < 0.001). The Google Fit® application demonstrated the highest reliability coefficient (ICC[2,1] = 0.93; p < 0.001; p = 0.37), compared to that of the GT3X® ActiGraph (ICC[2,1] = 0.32; p < 0.001; p < 0.001).
Conclusions:
The GT3X® ActiGraph underestimated the data and may not be the most appropriate device to estimate the stepping activity of stroke patients. The findings support the validity of a smartphone application in estimating the stepping activity of individuals with stroke, when worn on the paretic side.